TY - JOUR
T1 - Adaptive iterative dose reduction using 3D processing for reduced- and low-dose pulmonary CT
T2 - Comparison with standard-dose CT for image noise reduction and radiological findings
AU - Ohno, Yoshiharu
AU - Takenaka, Daisuke
AU - Kanda, Tomonori
AU - Yoshikawa, Takeshi
AU - Matsumoto, Sumiaki
AU - Sugihara, Naoki
AU - Sugimura, Kazuro
N1 - Funding Information:
1This research was supported by the MURST project ‘‘Metodi Variazionali ed Equazioni Differenziali non Lineari.’’
PY - 2012/10
Y1 - 2012/10
N2 - OBJECTIVE. The purpose of this study was to determine the utility of adaptive iterative dose reduction using 3D processing (AIDR 3D) for image noise reduction and assessment of radiologic findings obtained with reduced- and low-dose chest CT in patients with various pulmonary diseases. SUBJECTS AND METHODS. Chest CT examinations at three different tube current settings and using 16- and 64-MDCT scanners were performed for 37 patients. Standarddose (150 mAs) data were reconstructed as thin-section CT without AIDR 3D, and low-dose (25 mAs) and reduced-dose (50 mAs) data were reconstructed as thin-section CT without and with AIDR 3D. To compare image quality, image noises at all CT doses were quantitatively assessed by region of interest measurements. For comparison of radiologic finding assessments, likelihoods of occurrence of emphysema, ground-glass opacity, reticular opacity, bronchiectasis, honeycomb pattern, and nodules were evaluated on a 5-point scale. Then, image noise and agreements of radiologic findings between standard-dose CT and others were statistically evaluated. RESULTS. The image quality scores of reduced- and low-dose CT without AIDR 3D were significantly lower than those of both protocols with AIDR 3D and standard-dose CT (p < 0.05). All intermethod agreements for emphysema, ground-glass opacity, bronchiectasis, honeycomb pattern, and nodules, except for those observed on low-dose CT without AIDR 3D, were almost perfect (κ > 0.81). CONCLUSION. AIDR 3D is useful for image noise reduction and assessment of radiologic findings obtained with reduced- and low-dose CT for patients with various pulmonary diseases.
AB - OBJECTIVE. The purpose of this study was to determine the utility of adaptive iterative dose reduction using 3D processing (AIDR 3D) for image noise reduction and assessment of radiologic findings obtained with reduced- and low-dose chest CT in patients with various pulmonary diseases. SUBJECTS AND METHODS. Chest CT examinations at three different tube current settings and using 16- and 64-MDCT scanners were performed for 37 patients. Standarddose (150 mAs) data were reconstructed as thin-section CT without AIDR 3D, and low-dose (25 mAs) and reduced-dose (50 mAs) data were reconstructed as thin-section CT without and with AIDR 3D. To compare image quality, image noises at all CT doses were quantitatively assessed by region of interest measurements. For comparison of radiologic finding assessments, likelihoods of occurrence of emphysema, ground-glass opacity, reticular opacity, bronchiectasis, honeycomb pattern, and nodules were evaluated on a 5-point scale. Then, image noise and agreements of radiologic findings between standard-dose CT and others were statistically evaluated. RESULTS. The image quality scores of reduced- and low-dose CT without AIDR 3D were significantly lower than those of both protocols with AIDR 3D and standard-dose CT (p < 0.05). All intermethod agreements for emphysema, ground-glass opacity, bronchiectasis, honeycomb pattern, and nodules, except for those observed on low-dose CT without AIDR 3D, were almost perfect (κ > 0.81). CONCLUSION. AIDR 3D is useful for image noise reduction and assessment of radiologic findings obtained with reduced- and low-dose CT for patients with various pulmonary diseases.
UR - http://www.scopus.com/inward/record.url?scp=84866864966&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84866864966&partnerID=8YFLogxK
U2 - 10.2214/AJR.11.8275
DO - 10.2214/AJR.11.8275
M3 - Article
C2 - 22997397
AN - SCOPUS:84866864966
SN - 0361-803X
VL - 199
SP - W477-W485
JO - American Journal of Roentgenology
JF - American Journal of Roentgenology
IS - 4
ER -