Dose estimation with CTDI100, air in computed tomography

Masanao Kobayashi, Yumi Kataoka, Yoshihiro Ida, Yasuki Asada, Syoichi Suzuki, Kazuhiro Katada

Research output: Contribution to journalArticle

Abstract

Although the principal dosimetric quantity in computed tomography (CT) can be assessed using a pencil ionization chamber with an active length of 100 mm, standard CT dosimetry phantoms of polymethylmethacrylate (PMMA) , and plates of aluminum, most facilities do not possess the requisites. We present a practical method of estimating CTDI(100, c), CTDI(100, p) and the half-value layer (HVL) from CTDI(100, air), which is measured parallel with the axis of rotation of the scanner to free-in-air. The three data chosen for this method of estimation were as follows: 1) the relation of HVL to CTDI(100, air) per radiographic exposure (mAs); 2) the relation of HVL to CTDI(100, c) per CTDI(100, air); 3) the relation of HVL to CTDI(100, p) per CTDI(100, air). The data were based on the measured values of six CT scanners, so as to avoid dependence on the technical characteristics of a specific manufacturer. The estimated value has a possible maximum uncertainty of 20%, although this method of estimation is practical for dose assessment.

Original languageEnglish
Pages (from-to)1526-1532
Number of pages7
JournalNippon Hoshasen Gijutsu Gakkai zasshi
Volume60
Issue number11
DOIs
Publication statusPublished - 01-01-2004

Fingerprint

Air
Tomography
X-Ray Computed Tomography Scanners
Polymethyl Methacrylate
Aluminum
Uncertainty

All Science Journal Classification (ASJC) codes

  • Medicine(all)

Cite this

Kobayashi, Masanao ; Kataoka, Yumi ; Ida, Yoshihiro ; Asada, Yasuki ; Suzuki, Syoichi ; Katada, Kazuhiro. / Dose estimation with CTDI100, air in computed tomography. In: Nippon Hoshasen Gijutsu Gakkai zasshi. 2004 ; Vol. 60, No. 11. pp. 1526-1532.
@article{38dde22fb76a421c85f6e1cced260b93,
title = "Dose estimation with CTDI100, air in computed tomography",
abstract = "Although the principal dosimetric quantity in computed tomography (CT) can be assessed using a pencil ionization chamber with an active length of 100 mm, standard CT dosimetry phantoms of polymethylmethacrylate (PMMA) , and plates of aluminum, most facilities do not possess the requisites. We present a practical method of estimating CTDI(100, c), CTDI(100, p) and the half-value layer (HVL) from CTDI(100, air), which is measured parallel with the axis of rotation of the scanner to free-in-air. The three data chosen for this method of estimation were as follows: 1) the relation of HVL to CTDI(100, air) per radiographic exposure (mAs); 2) the relation of HVL to CTDI(100, c) per CTDI(100, air); 3) the relation of HVL to CTDI(100, p) per CTDI(100, air). The data were based on the measured values of six CT scanners, so as to avoid dependence on the technical characteristics of a specific manufacturer. The estimated value has a possible maximum uncertainty of 20{\%}, although this method of estimation is practical for dose assessment.",
author = "Masanao Kobayashi and Yumi Kataoka and Yoshihiro Ida and Yasuki Asada and Syoichi Suzuki and Kazuhiro Katada",
year = "2004",
month = "1",
day = "1",
doi = "10.6009/jjrt.KJ00003326576",
language = "English",
volume = "60",
pages = "1526--1532",
journal = "Nippon Hoshasen Gijutsu Gakkai zasshi",
issn = "0369-4305",
publisher = "Nihon Hoshasen Gijutsu Gakkai/Japanese Society of Radiological Technology",
number = "11",

}

Kobayashi, M, Kataoka, Y, Ida, Y, Asada, Y, Suzuki, S & Katada, K 2004, 'Dose estimation with CTDI100, air in computed tomography', Nippon Hoshasen Gijutsu Gakkai zasshi, vol. 60, no. 11, pp. 1526-1532. https://doi.org/10.6009/jjrt.KJ00003326576

Dose estimation with CTDI100, air in computed tomography. / Kobayashi, Masanao; Kataoka, Yumi; Ida, Yoshihiro; Asada, Yasuki; Suzuki, Syoichi; Katada, Kazuhiro.

In: Nippon Hoshasen Gijutsu Gakkai zasshi, Vol. 60, No. 11, 01.01.2004, p. 1526-1532.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Dose estimation with CTDI100, air in computed tomography

AU - Kobayashi, Masanao

AU - Kataoka, Yumi

AU - Ida, Yoshihiro

AU - Asada, Yasuki

AU - Suzuki, Syoichi

AU - Katada, Kazuhiro

PY - 2004/1/1

Y1 - 2004/1/1

N2 - Although the principal dosimetric quantity in computed tomography (CT) can be assessed using a pencil ionization chamber with an active length of 100 mm, standard CT dosimetry phantoms of polymethylmethacrylate (PMMA) , and plates of aluminum, most facilities do not possess the requisites. We present a practical method of estimating CTDI(100, c), CTDI(100, p) and the half-value layer (HVL) from CTDI(100, air), which is measured parallel with the axis of rotation of the scanner to free-in-air. The three data chosen for this method of estimation were as follows: 1) the relation of HVL to CTDI(100, air) per radiographic exposure (mAs); 2) the relation of HVL to CTDI(100, c) per CTDI(100, air); 3) the relation of HVL to CTDI(100, p) per CTDI(100, air). The data were based on the measured values of six CT scanners, so as to avoid dependence on the technical characteristics of a specific manufacturer. The estimated value has a possible maximum uncertainty of 20%, although this method of estimation is practical for dose assessment.

AB - Although the principal dosimetric quantity in computed tomography (CT) can be assessed using a pencil ionization chamber with an active length of 100 mm, standard CT dosimetry phantoms of polymethylmethacrylate (PMMA) , and plates of aluminum, most facilities do not possess the requisites. We present a practical method of estimating CTDI(100, c), CTDI(100, p) and the half-value layer (HVL) from CTDI(100, air), which is measured parallel with the axis of rotation of the scanner to free-in-air. The three data chosen for this method of estimation were as follows: 1) the relation of HVL to CTDI(100, air) per radiographic exposure (mAs); 2) the relation of HVL to CTDI(100, c) per CTDI(100, air); 3) the relation of HVL to CTDI(100, p) per CTDI(100, air). The data were based on the measured values of six CT scanners, so as to avoid dependence on the technical characteristics of a specific manufacturer. The estimated value has a possible maximum uncertainty of 20%, although this method of estimation is practical for dose assessment.

UR - http://www.scopus.com/inward/record.url?scp=16644362586&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=16644362586&partnerID=8YFLogxK

U2 - 10.6009/jjrt.KJ00003326576

DO - 10.6009/jjrt.KJ00003326576

M3 - Article

C2 - 15568004

AN - SCOPUS:16644362586

VL - 60

SP - 1526

EP - 1532

JO - Nippon Hoshasen Gijutsu Gakkai zasshi

JF - Nippon Hoshasen Gijutsu Gakkai zasshi

SN - 0369-4305

IS - 11

ER -