TY - JOUR
T1 - Early experience with low-prime (99 ml) extracorporeal membrane oxygenation support in children
AU - Yamasaki, Yasumori
AU - Hayashi, Teruyuki
AU - Nakatani, Takeshi
AU - Yotsuida, Hideki
AU - Nishigaki, Takayuki
AU - Takahashi, Yuzo
AU - Inamori, Shuji
AU - Kagisaki, Kuji
AU - Hagino, H.
AU - Ishizaka, T.
AU - Yagihara, Toshikatsu
PY - 2006/1
Y1 - 2006/1
N2 - Quick setup is mandatory for cardiopulmonary resuscitation using an extracorporeal membrane oxygenation (ECMO) assist device. Our conventional ECMO circuit for pediatric patients consists of a centrifugal pump (CX-HP) and membrane oxygenator (CX10H). Because of the large priming volume (260 ml), the circuit had to be primed with donor blood and required 30 minutes for setup. We started to use a low-prime ECMO with small centrifugal pump (HPM-15) and membrane oxygenator (MENOX Alpha Cube) for induction of ECMO beginning in 2000. The priming volume of this low-prime circuit is only 99 ml. The circuit can be primed without donor blood, even in the small patient, and requires only 10 minutes to set up. We review our experiences with cardiopulmonary resuscitation for sudden cardiopulmonary collapse in pediatric patients, including postcardiotomy patients. From 1997 to 2000, 23 patients underwent ECMO support with a conventional circuit (group A). From 2000 to 2004, we used low-prime circuit for induction of ECMO in 12 patients (group B). After the induction of ECMO with low-prime circuit, ECMO was converted to conventional heparin-bonded circuit for the longer support. The results suggested that the quick induction of ECMO with low-prime circuit has significant advantages in cardiopulmonary support in pediatric patients.
AB - Quick setup is mandatory for cardiopulmonary resuscitation using an extracorporeal membrane oxygenation (ECMO) assist device. Our conventional ECMO circuit for pediatric patients consists of a centrifugal pump (CX-HP) and membrane oxygenator (CX10H). Because of the large priming volume (260 ml), the circuit had to be primed with donor blood and required 30 minutes for setup. We started to use a low-prime ECMO with small centrifugal pump (HPM-15) and membrane oxygenator (MENOX Alpha Cube) for induction of ECMO beginning in 2000. The priming volume of this low-prime circuit is only 99 ml. The circuit can be primed without donor blood, even in the small patient, and requires only 10 minutes to set up. We review our experiences with cardiopulmonary resuscitation for sudden cardiopulmonary collapse in pediatric patients, including postcardiotomy patients. From 1997 to 2000, 23 patients underwent ECMO support with a conventional circuit (group A). From 2000 to 2004, we used low-prime circuit for induction of ECMO in 12 patients (group B). After the induction of ECMO with low-prime circuit, ECMO was converted to conventional heparin-bonded circuit for the longer support. The results suggested that the quick induction of ECMO with low-prime circuit has significant advantages in cardiopulmonary support in pediatric patients.
UR - http://www.scopus.com/inward/record.url?scp=33645738450&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=33645738450&partnerID=8YFLogxK
U2 - 10.1097/01.mat.0000178960.76823.c7
DO - 10.1097/01.mat.0000178960.76823.c7
M3 - Article
C2 - 16436900
AN - SCOPUS:33645738450
SN - 1058-2916
VL - 52
SP - 110
EP - 114
JO - ASAIO Journal
JF - ASAIO Journal
IS - 1
ER -