Electroretinographic Monitoring of Retinal Function During Eye Surgery

Yozo Miyake, Katsuya Yagasaki, Masayuki Horiguchi

Research output: Contribution to journalArticlepeer-review

31 Citations (Scopus)

Abstract

We performed a pilot study of electroretinographic monitoring of retinal function during eye surgery performed with local anesthesia. A contact lens with a built-in light-emitting diode was sterilized and used as both a stimulus source and a recording electrode for 30-Hz flicker electroretinograms. Each recording required 7 seconds, during which the operating light (but not the room light) was switched off. Electroretinograms recorded during surgery on 10 patients with minimal retinal abnormality were evaluated for the effects of light adaptation caused by regular exposure to an operating light source or operating microscope. The fluctuations of the amplitude and peak time were small, indicating that this technique can be used to monitor retinal function during eye surgery. Some results during retinal and vitreous surgeries are shown.

Original languageEnglish
Pages (from-to)1123-1126
Number of pages4
JournalArchives of Ophthalmology
Volume109
Issue number8
DOIs
Publication statusPublished - 08-1991
Externally publishedYes

All Science Journal Classification (ASJC) codes

  • Ophthalmology

Fingerprint

Dive into the research topics of 'Electroretinographic Monitoring of Retinal Function During Eye Surgery'. Together they form a unique fingerprint.