Evaluation of geometric efficiency and radiation exposure in z-axis for volume scan

Masanao Kobayashi, Kichiro Koshida, Shoichi Suzuki, Kazuhiro Katada

Research output: Contribution to journalArticle

6 Citations (Scopus)

Abstract

The International Electrotechnical Commission (IEC) recommended that a display be provided on the operating console of medical computed tomography (CT) scanners when the geometric efficiency (GE) in the z-axis is 70 % or less. Taking into account the increase in the number of detector rows and the changes in the scanning method, the present study was conducted to review the GE and to evaluate the exposure dose in the z-axis direction. The GE for a single-rotation scan was calculated in accordance with IEC-606024 Ed. 2. The findings of the present study suggest that 320-row area detector computed tomography volume scan (single axial scan with no table movement in which it just happened to use a very large beam width) extends 18.8 mm beyond the nominal beam collimation, even when the GE is >70 %. However, compared with the helical scan, which is the present mainstream of CT examinations, the dose beyond the scan length is narrower in volume scan for a limited scan length of 160 mm or less.

Original languageEnglish
Article numberncq283
Pages (from-to)63-68
Number of pages6
JournalRadiation Protection Dosimetry
Volume143
Issue number1
DOIs
Publication statusPublished - 01-01-2011

Fingerprint

X-Ray Computed Tomography Scanners
Cone-Beam Computed Tomography
radiation dosage
Tomography
evaluation
tomography
consoles
dosage
detectors
collimation
scanners
examination
scanning
Direction compound
Radiation Exposure

All Science Journal Classification (ASJC) codes

  • Radiology Nuclear Medicine and imaging
  • Radiological and Ultrasound Technology
  • Radiation
  • Public Health, Environmental and Occupational Health

Cite this

Kobayashi, Masanao ; Koshida, Kichiro ; Suzuki, Shoichi ; Katada, Kazuhiro. / Evaluation of geometric efficiency and radiation exposure in z-axis for volume scan. In: Radiation Protection Dosimetry. 2011 ; Vol. 143, No. 1. pp. 63-68.
@article{cdc8e0108d4e4fff86dcdaa045c2a3d5,
title = "Evaluation of geometric efficiency and radiation exposure in z-axis for volume scan",
abstract = "The International Electrotechnical Commission (IEC) recommended that a display be provided on the operating console of medical computed tomography (CT) scanners when the geometric efficiency (GE) in the z-axis is 70 {\%} or less. Taking into account the increase in the number of detector rows and the changes in the scanning method, the present study was conducted to review the GE and to evaluate the exposure dose in the z-axis direction. The GE for a single-rotation scan was calculated in accordance with IEC-606024 Ed. 2. The findings of the present study suggest that 320-row area detector computed tomography volume scan (single axial scan with no table movement in which it just happened to use a very large beam width) extends 18.8 mm beyond the nominal beam collimation, even when the GE is >70 {\%}. However, compared with the helical scan, which is the present mainstream of CT examinations, the dose beyond the scan length is narrower in volume scan for a limited scan length of 160 mm or less.",
author = "Masanao Kobayashi and Kichiro Koshida and Shoichi Suzuki and Kazuhiro Katada",
year = "2011",
month = "1",
day = "1",
doi = "10.1093/rpd/ncq283",
language = "English",
volume = "143",
pages = "63--68",
journal = "Radiation Protection Dosimetry",
issn = "0144-8420",
publisher = "Oxford University Press",
number = "1",

}

Evaluation of geometric efficiency and radiation exposure in z-axis for volume scan. / Kobayashi, Masanao; Koshida, Kichiro; Suzuki, Shoichi; Katada, Kazuhiro.

In: Radiation Protection Dosimetry, Vol. 143, No. 1, ncq283, 01.01.2011, p. 63-68.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Evaluation of geometric efficiency and radiation exposure in z-axis for volume scan

AU - Kobayashi, Masanao

AU - Koshida, Kichiro

AU - Suzuki, Shoichi

AU - Katada, Kazuhiro

PY - 2011/1/1

Y1 - 2011/1/1

N2 - The International Electrotechnical Commission (IEC) recommended that a display be provided on the operating console of medical computed tomography (CT) scanners when the geometric efficiency (GE) in the z-axis is 70 % or less. Taking into account the increase in the number of detector rows and the changes in the scanning method, the present study was conducted to review the GE and to evaluate the exposure dose in the z-axis direction. The GE for a single-rotation scan was calculated in accordance with IEC-606024 Ed. 2. The findings of the present study suggest that 320-row area detector computed tomography volume scan (single axial scan with no table movement in which it just happened to use a very large beam width) extends 18.8 mm beyond the nominal beam collimation, even when the GE is >70 %. However, compared with the helical scan, which is the present mainstream of CT examinations, the dose beyond the scan length is narrower in volume scan for a limited scan length of 160 mm or less.

AB - The International Electrotechnical Commission (IEC) recommended that a display be provided on the operating console of medical computed tomography (CT) scanners when the geometric efficiency (GE) in the z-axis is 70 % or less. Taking into account the increase in the number of detector rows and the changes in the scanning method, the present study was conducted to review the GE and to evaluate the exposure dose in the z-axis direction. The GE for a single-rotation scan was calculated in accordance with IEC-606024 Ed. 2. The findings of the present study suggest that 320-row area detector computed tomography volume scan (single axial scan with no table movement in which it just happened to use a very large beam width) extends 18.8 mm beyond the nominal beam collimation, even when the GE is >70 %. However, compared with the helical scan, which is the present mainstream of CT examinations, the dose beyond the scan length is narrower in volume scan for a limited scan length of 160 mm or less.

UR - http://www.scopus.com/inward/record.url?scp=78651454753&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=78651454753&partnerID=8YFLogxK

U2 - 10.1093/rpd/ncq283

DO - 10.1093/rpd/ncq283

M3 - Article

C2 - 20926511

AN - SCOPUS:78651454753

VL - 143

SP - 63

EP - 68

JO - Radiation Protection Dosimetry

JF - Radiation Protection Dosimetry

SN - 0144-8420

IS - 1

M1 - ncq283

ER -