TY - JOUR
T1 - Evaluation of radiophotoluminescent glass dosimeter response for therapeutic spot scanning proton beam
T2 - suggestion of linear energy transfer-based correction
AU - Nagata, Junya
AU - Yasui, Keisuke
AU - Omachi, Chihiro
AU - Toshiyuki, Toshito
AU - Shimizu, Hidetoshi
AU - Aoyama, Takahiro
AU - Hayashi, Naoki
N1 - Publisher Copyright:
© 2021 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine
PY - 2021/8
Y1 - 2021/8
N2 - A radiophotoluminescent glass dosimeter (RGD) is used for a postal audit of a photon beam because of its various excellent characteristics. However, it has not been used for scanning proton beams because its response characteristics have not been verified. In this study, the response of RGD to scanning protons was investigated to develop a dosimetry protocol using the linear energy transfer (LET)-based correction factor. The responses of RGD to four maximum-range-energy-pattern proton beams were verified by comparing it with ionization chamber (IC) dosimetry. The LET at each measurement depth was calculated via Monte Carlo (MC) simulation. The LET correction factor ((Formula presented.)) was the ratio between the uncorrected RGD dose ((Formula presented.)) and the IC dose at each measurement depth. (Formula presented.) can be represented as a function of LET using the following equation: (Formula presented.). (Formula presented.) showed a linear under-response with increasing LET, and the maximum dose difference between the IC dose and (Formula presented.) was 15.2% at an LET of 6.07 keV/μm. The LET-based correction dose ((Formula presented.)) conformed within 3.6% of the IC dose. The mean dose difference (±SD) of (Formula presented.) and (Formula presented.) was –2.5 ± 6.9% and 0.0 ± 1.6%, respectively. To achieve accurate dose verification for scanning proton beams using RGD, we derived a linear regression equation based on LET. The results show that with appropriate LET correction, RGD can be used for dose verification of scanning proton beams.
AB - A radiophotoluminescent glass dosimeter (RGD) is used for a postal audit of a photon beam because of its various excellent characteristics. However, it has not been used for scanning proton beams because its response characteristics have not been verified. In this study, the response of RGD to scanning protons was investigated to develop a dosimetry protocol using the linear energy transfer (LET)-based correction factor. The responses of RGD to four maximum-range-energy-pattern proton beams were verified by comparing it with ionization chamber (IC) dosimetry. The LET at each measurement depth was calculated via Monte Carlo (MC) simulation. The LET correction factor ((Formula presented.)) was the ratio between the uncorrected RGD dose ((Formula presented.)) and the IC dose at each measurement depth. (Formula presented.) can be represented as a function of LET using the following equation: (Formula presented.). (Formula presented.) showed a linear under-response with increasing LET, and the maximum dose difference between the IC dose and (Formula presented.) was 15.2% at an LET of 6.07 keV/μm. The LET-based correction dose ((Formula presented.)) conformed within 3.6% of the IC dose. The mean dose difference (±SD) of (Formula presented.) and (Formula presented.) was –2.5 ± 6.9% and 0.0 ± 1.6%, respectively. To achieve accurate dose verification for scanning proton beams using RGD, we derived a linear regression equation based on LET. The results show that with appropriate LET correction, RGD can be used for dose verification of scanning proton beams.
UR - http://www.scopus.com/inward/record.url?scp=85111926898&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85111926898&partnerID=8YFLogxK
U2 - 10.1002/acm2.13378
DO - 10.1002/acm2.13378
M3 - Article
C2 - 34339583
AN - SCOPUS:85111926898
SN - 1526-9914
VL - 22
SP - 265
EP - 272
JO - Journal of applied clinical medical physics
JF - Journal of applied clinical medical physics
IS - 8
ER -