Skip to main navigation Skip to search Skip to main content

Fully Automated Rapid Immunofludic Device by Use of Microfibrous Reactor for Human Saliva Sample

  • Yecan Wang
  • , Kaito Maehara
  • , Shigenobu Mitsuzawa
  • , Satoru Shinkawa
  • , Toshihiro Kasama
  • , Ryo Miyake
  • , Madoka Takai

Research output: Chapter in Book/Report/Conference proceedingConference contribution

Abstract

Given the growing need for rapid measurement for health monitoring, small sized automated analysis devices from human sample directly are required. We developed fully automated immunofluidic device for rapid biomarker detection from human saliva sample directly. The total system detects human cortisol in 15 minutes with high sensitivity (0.5∼ 30 ng/mL) and a limit detection of 0.36 ng/mL, integrating pretreatment, immunoreaction, and analysis efficiently.

Original languageEnglish
Title of host publication2024 IEEE Sensors, SENSORS 2024 - Conference Proceedings
PublisherInstitute of Electrical and Electronics Engineers Inc.
ISBN (Electronic)9798350363517
DOIs
Publication statusPublished - 2024
Externally publishedYes
Event2024 IEEE Sensors, SENSORS 2024 - Kobe, Japan
Duration: 20-10-202423-10-2024

Publication series

NameProceedings of IEEE Sensors
ISSN (Print)1930-0395
ISSN (Electronic)2168-9229

Conference

Conference2024 IEEE Sensors, SENSORS 2024
Country/TerritoryJapan
CityKobe
Period20-10-2423-10-24

UN SDGs

This output contributes to the following UN Sustainable Development Goals (SDGs)

  1. SDG 3 - Good Health and Well-being
    SDG 3 Good Health and Well-being

All Science Journal Classification (ASJC) codes

  • Electrical and Electronic Engineering

Fingerprint

Dive into the research topics of 'Fully Automated Rapid Immunofludic Device by Use of Microfibrous Reactor for Human Saliva Sample'. Together they form a unique fingerprint.

Cite this