TY - JOUR
T1 - Investigation of measurement accuracy of factors used for detective quantum efficiency measurement in digital radiography.
AU - Kunitomo, Hiroshi
AU - Koyama, Shuji
AU - Higashide, Ryo
AU - Ichikawa, Katsuhiro
AU - Hattori, Masumi
AU - Okada, Yoko
AU - Hayashi, Norio
AU - Sawada, Michito
PY - 2014/7
Y1 - 2014/7
N2 - In the detective quantum efficiency (DQE) evaluation of detectors for digital radiography (DR) systems, physical image quality indices such as modulation transfer function (MTF) and normalized noise power spectrum (NNPS) need to be accurately measured to obtain highly accurate DQE evaluations. However, there is a risk of errors in these measurements. In this study, we focused on error factors that should be considered in measurements using clinical DR systems. We compared the incident photon numbers indicated in IEC 62220-1 with those estimated using a Monte Carlo simulation based on X-ray energy spectra measured employing four DR systems. For NNPS, influences of X-ray intensity non-uniformity, tube voltage and aluminum purity were investigated. The effects of geometric magnifications on MTF accuracy were also examined using a tungsten edge plate at distances of 50, 100 and 150 mm from the detector surface at a source-image receptor distance of 2000 mm. The photon numbers in IEC 62220-1 coincided with our estimates of values, with error rates below 2.5%. Tube voltage errors of approximately ±5 kV caused NNPS errors of within 1.0%. The X-ray intensity non-uniformity caused NNPS errors of up to 2.0% at the anode side. Aluminum purity did not affect the measurement accuracy. The maximum MTF reductions caused by geometric magnifications were 3.67% for 1.0-mm X-ray focus and 1.83% for 0.6-mm X-ray focus.
AB - In the detective quantum efficiency (DQE) evaluation of detectors for digital radiography (DR) systems, physical image quality indices such as modulation transfer function (MTF) and normalized noise power spectrum (NNPS) need to be accurately measured to obtain highly accurate DQE evaluations. However, there is a risk of errors in these measurements. In this study, we focused on error factors that should be considered in measurements using clinical DR systems. We compared the incident photon numbers indicated in IEC 62220-1 with those estimated using a Monte Carlo simulation based on X-ray energy spectra measured employing four DR systems. For NNPS, influences of X-ray intensity non-uniformity, tube voltage and aluminum purity were investigated. The effects of geometric magnifications on MTF accuracy were also examined using a tungsten edge plate at distances of 50, 100 and 150 mm from the detector surface at a source-image receptor distance of 2000 mm. The photon numbers in IEC 62220-1 coincided with our estimates of values, with error rates below 2.5%. Tube voltage errors of approximately ±5 kV caused NNPS errors of within 1.0%. The X-ray intensity non-uniformity caused NNPS errors of up to 2.0% at the anode side. Aluminum purity did not affect the measurement accuracy. The maximum MTF reductions caused by geometric magnifications were 3.67% for 1.0-mm X-ray focus and 1.83% for 0.6-mm X-ray focus.
UR - http://www.scopus.com/inward/record.url?scp=84906926612&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84906926612&partnerID=8YFLogxK
U2 - 10.6009/jjrt.2014_JSRT_70.7.653
DO - 10.6009/jjrt.2014_JSRT_70.7.653
M3 - Article
C2 - 25055945
AN - SCOPUS:84906926612
SN - 0369-4305
VL - 70
SP - 653
EP - 661
JO - Nihon Hoshasen Gijutsu Gakkai zasshi
JF - Nihon Hoshasen Gijutsu Gakkai zasshi
IS - 7
ER -