TY - JOUR
T1 - Optimization of energy windows to calculate quantitative X-ray images using an energy-resolving photon-counting detector
T2 - A simulation study
AU - Nishigami, Rina
AU - Kobayashi, Daiki
AU - Kimoto, Natsumi
AU - Asahara, Takashi
AU - Maeda, Tatsuya
AU - Haba, Tomonobu
AU - Kanazawa, Yuki
AU - Yamamoto, Shuichiro
AU - Hayashi, Hiroaki
N1 - Publisher Copyright:
© 2024 Elsevier Ltd
PY - 2025/4
Y1 - 2025/4
N2 - Purpose: Energy-resolving photon-counting detectors (ERPCDs) are expected to enable novel functional diagnosis using quantitative X-ray images such as effective atomic number (Zeff) images. In this simulation study, we sought to optimize the settings related to tube voltage and energy windows. Methods: We assumed a virtual phantom composed of polymethylmethacrylate (PMMA, Zeff = 6.5) and aluminum (Zeff = 13.0), and simulated the conventional energy integrating detector (EID) image and Zeff images obtained by the ERPCD. The investigational phantom is composed of elements with ρt ranging from 0.1 to 80 g/cm2. In order to perform optimization using a quantitative index, we defined a system performance function (SPF) that takes into account the contributions of the contrast to noise ratio (CNR) of the EID image and the uncertainty (δZeff) of the Zeff image. The tube voltage was varied to be 60, 90, and 120 kV, and the variable that determines the separation energy between the middle and high energy windows was changed from 35 to 115 keV with 5 keV interval. In addition, preclinical images of the digital phantoms based on computed tomography (CT) images were created for demonstration. Results: We were able to determine the imaging conditions that yielded the better image quality for tube voltages of 60, 90, and 120 kV. Among these, the 120 kV condition (20-30-50-120 keV) showed the smallest SPF value, and was therefore adopted as the optimal condition. Furthermore, high-quality EID and Zeff images of preclinical phantoms were obtained under the optimal condition. Conclusions: We determined the optimal condition suitable for deriving both qualitative and quantitative images using ERPCDs.
AB - Purpose: Energy-resolving photon-counting detectors (ERPCDs) are expected to enable novel functional diagnosis using quantitative X-ray images such as effective atomic number (Zeff) images. In this simulation study, we sought to optimize the settings related to tube voltage and energy windows. Methods: We assumed a virtual phantom composed of polymethylmethacrylate (PMMA, Zeff = 6.5) and aluminum (Zeff = 13.0), and simulated the conventional energy integrating detector (EID) image and Zeff images obtained by the ERPCD. The investigational phantom is composed of elements with ρt ranging from 0.1 to 80 g/cm2. In order to perform optimization using a quantitative index, we defined a system performance function (SPF) that takes into account the contributions of the contrast to noise ratio (CNR) of the EID image and the uncertainty (δZeff) of the Zeff image. The tube voltage was varied to be 60, 90, and 120 kV, and the variable that determines the separation energy between the middle and high energy windows was changed from 35 to 115 keV with 5 keV interval. In addition, preclinical images of the digital phantoms based on computed tomography (CT) images were created for demonstration. Results: We were able to determine the imaging conditions that yielded the better image quality for tube voltages of 60, 90, and 120 kV. Among these, the 120 kV condition (20-30-50-120 keV) showed the smallest SPF value, and was therefore adopted as the optimal condition. Furthermore, high-quality EID and Zeff images of preclinical phantoms were obtained under the optimal condition. Conclusions: We determined the optimal condition suitable for deriving both qualitative and quantitative images using ERPCDs.
KW - Medical X-ray diagnosis
KW - Optimization of imaging condition
KW - Photon counting imaging detector
KW - Quantitative X-ray image
UR - http://www.scopus.com/inward/record.url?scp=85211167968&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85211167968&partnerID=8YFLogxK
U2 - 10.1016/j.radphyschem.2024.112460
DO - 10.1016/j.radphyschem.2024.112460
M3 - Article
AN - SCOPUS:85211167968
SN - 0969-806X
VL - 229
JO - Radiation Physics and Chemistry
JF - Radiation Physics and Chemistry
M1 - 112460
ER -