TY - JOUR
T1 - Optimizing LM-DRAMA parameters and non-local means filtering to improve small-lesion detectability in SiPM-based TOF breast PET
AU - Shiiba, Takuro
AU - Katakami, Hana
AU - Naito, Aiko
AU - Takamura, Maki
AU - Ishiguro, Masanobu
AU - Watanabe, Masanori
AU - Uno, Masaki
N1 - Publisher Copyright:
© Australasian College of Physical Scientists and Engineers in Medicine 2025.
PY - 2025/9
Y1 - 2025/9
N2 - This study aimed to optimize image reconstruction parameters for a dedicated time-of-flight (TOF) breast positron emission tomography (PET) system equipped with silicon photomultipliers (SiPMs) that maximize lesion detectability while minimizing image noise. A cylindrical phantom containing four hot spheres (3–10 mm diameter) was scanned at sphere-to-background ratios of 4:1, 6:1, and 8:1. All data were reconstructed using a 3D list-mode dynamic row-action maximum likelihood algorithm with β values of 10–200, followed by non-local means (NLM) filtering at intensities of 0.5–2.0 or no filtering. Image quality was evaluated using background coefficient of variation (COVBG), contrast recovery coefficient (CRC), and detectability index (DI) for the 3 mm sphere. As β increased, CRC and DI improved, particularly for smaller spheres and higher SBRs; however, background noise also rose. Applying the NLM filter reduced COVBG, especially when increasing the filter intensity from 0.5 to 1.0, although noise reduction gains plateaued at intensities above 1.0. Optimal trade-offs in lesion detectability and noise were observed at moderate β (50–100) with NLM intensities of 1.0–1.5, yielding higher CRC and DI without excessive background noise or blurring effects. A balanced approach to β and NLM filtering substantially enhances small-lesion visibility in SiPM-based TOF-dedicated breast PET imaging. These findings offer a practical framework for parameter selection, supporting better lesion detectability and advancing breast cancer diagnostics through more sensitive PET protocols.
AB - This study aimed to optimize image reconstruction parameters for a dedicated time-of-flight (TOF) breast positron emission tomography (PET) system equipped with silicon photomultipliers (SiPMs) that maximize lesion detectability while minimizing image noise. A cylindrical phantom containing four hot spheres (3–10 mm diameter) was scanned at sphere-to-background ratios of 4:1, 6:1, and 8:1. All data were reconstructed using a 3D list-mode dynamic row-action maximum likelihood algorithm with β values of 10–200, followed by non-local means (NLM) filtering at intensities of 0.5–2.0 or no filtering. Image quality was evaluated using background coefficient of variation (COVBG), contrast recovery coefficient (CRC), and detectability index (DI) for the 3 mm sphere. As β increased, CRC and DI improved, particularly for smaller spheres and higher SBRs; however, background noise also rose. Applying the NLM filter reduced COVBG, especially when increasing the filter intensity from 0.5 to 1.0, although noise reduction gains plateaued at intensities above 1.0. Optimal trade-offs in lesion detectability and noise were observed at moderate β (50–100) with NLM intensities of 1.0–1.5, yielding higher CRC and DI without excessive background noise or blurring effects. A balanced approach to β and NLM filtering substantially enhances small-lesion visibility in SiPM-based TOF-dedicated breast PET imaging. These findings offer a practical framework for parameter selection, supporting better lesion detectability and advancing breast cancer diagnostics through more sensitive PET protocols.
KW - Dedicated breast positron emission tomography
KW - Dynamic row-action maximum likelihood algorithm
KW - Non-local means filter
KW - Silicon photomultiplier
UR - https://www.scopus.com/pages/publications/105011359833
UR - https://www.scopus.com/pages/publications/105011359833#tab=citedBy
U2 - 10.1007/s13246-025-01598-y
DO - 10.1007/s13246-025-01598-y
M3 - Article
C2 - 40694234
AN - SCOPUS:105011359833
SN - 2662-4729
VL - 48
SP - 1453
EP - 1461
JO - Physical and Engineering Sciences in Medicine
JF - Physical and Engineering Sciences in Medicine
IS - 3
ER -