T1 - Organ-Specific Positron Emission Tomography Scanners for Breast Imaging
T2 - Comparison between the Performances of Prior and Novel Models
AU - Satoh, Yoko
AU - Hanaoka, Kohei
AU - Ikegawa, Chihiro
AU - Imai, Masamichi
AU - Watanabe, Shota
AU - Morimoto-Ishikawa, Daisuke
AU - Onishi, Hiroshi
AU - Ito, Toshikazu
AU - Komoike, Yoshifumi
AU - Ishii, Kazunari
N1 - Publisher Copyright:
© 2023 by the authors.
PY - 2023/3
Y1 - 2023/3
N2 - The performances of photomultiplier tube (PMT)-based dedicated breast positron emission tomography (PET) and silicon photomultiplier tube (SiPM)-based time-of-flight (TOF) PET, which is applicable not only to breast imaging but also to head imaging, were compared using a phantom study. A cylindrical phantom containing four spheres (3–10 mm in diameter) filled with 18F-FDG at two signal-to-background ratios (SBRs), 4:1 and 8:1, was scanned. The phantom images, which were reconstructed using three-dimensional list-mode dynamic row-action maximum likelihood algorithm with various β-values and post-smoothing filters, were visually and quantitatively compared. Visual evaluation showed that the 3 mm sphere was more clearly visualized with higher β and smaller post-filters, while the background was noisier; SiPM-based TOF-PET was superior to PMT-based dbPET in sharpness, smoothness, and detectability, although the background was noisier at the SBR of 8:1. Quantitative evaluation revealed that the detection index (DI) and recovery coefficient (CRC) of SiPM-based TOF-PET images were higher than those of PMT-based PET images, despite a higher background coefficient of variation (CVBG). The two organ-specific PET systems showed that a 3 mm lesion in the breast could be visualized at the center of the detector, and there was less noise in the SiPM-based TOF-PET image.
AB - The performances of photomultiplier tube (PMT)-based dedicated breast positron emission tomography (PET) and silicon photomultiplier tube (SiPM)-based time-of-flight (TOF) PET, which is applicable not only to breast imaging but also to head imaging, were compared using a phantom study. A cylindrical phantom containing four spheres (3–10 mm in diameter) filled with 18F-FDG at two signal-to-background ratios (SBRs), 4:1 and 8:1, was scanned. The phantom images, which were reconstructed using three-dimensional list-mode dynamic row-action maximum likelihood algorithm with various β-values and post-smoothing filters, were visually and quantitatively compared. Visual evaluation showed that the 3 mm sphere was more clearly visualized with higher β and smaller post-filters, while the background was noisier; SiPM-based TOF-PET was superior to PMT-based dbPET in sharpness, smoothness, and detectability, although the background was noisier at the SBR of 8:1. Quantitative evaluation revealed that the detection index (DI) and recovery coefficient (CRC) of SiPM-based TOF-PET images were higher than those of PMT-based PET images, despite a higher background coefficient of variation (CVBG). The two organ-specific PET systems showed that a 3 mm lesion in the breast could be visualized at the center of the detector, and there was less noise in the SiPM-based TOF-PET image.
UR - http://www.scopus.com/inward/record.url?scp=85151390724&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85151390724&partnerID=8YFLogxK
U2 - 10.3390/diagnostics13061079
DO - 10.3390/diagnostics13061079
M3 - Article
AN - SCOPUS:85151390724
SN - 2075-4418
VL - 13
JO - Diagnostics
JF - Diagnostics
IS - 6
M1 - 1079
ER -