Persistent contrast-filling in the woven endobridge device three months after its implantation for cerebral aneurysm: Incidence, predictive factors, and outcome

Research output: Contribution to journalArticlepeer-review

Abstract

Objective: The Woven EndoBridge (WEB) was developed to treat wide-neck bifurcation intracranial aneurysms. Occasionally, persistent contrast-filling has been observed in the WEB after treatment. The purpose of our study was to investigate its incidence, predictive factors, and clinical impact. Methods: All patients treated with the WEB between January 2021 and September 2021 at our institute were reviewed. Age, gender, antiplatelet therapy, and angioarchitecture were compared between the persistent-filling group and the no-filling group at the three-month follow-up angiography. Results: We included 20 patients with 20 unruptured aneurysms. Ten of the 20 intracranial aneurysms (50 %) showed contrast-filling in the WEB after three months. Two of the 10 had contrast not only inside, but around the device. Statistically significant differences were observed between the persistent-filling group and the no-filling group in neck size (median: 4.5 mm vs. 3.8 mm), deviation of the aneurysm axis from the inlet flow line where the orifice of the bifurcated arteries overlaps (mean: 15.1° vs. 33.0°), and postoperative dual antiplatelet therapy (DAPT) for at least a month (90 % vs. 20 %). One case had additional coil embolization six months after the WEB implantation. Including this case, one year after the treatment or the re-treatment, the filling in the device had disappeared in nine of 10 cases. No bleeding was observed during the mean follow-up period of 24 months. Conclusion: Persistent contrast-filling was associated with postoperative DAPT for at least a month, a wide neck, and less deviation of the aneurysm axis from the inlet flow line. If the contrast-filling is only within the WEB and not between the aneurysmal wall and the WEB, we are not concerned. To further assess the clinical impact of this phenomenon, long-term follow-up will be needed.

Original languageEnglish
Article number107837
JournalClinical Neurology and Neurosurgery
Volume231
DOIs
Publication statusPublished - 08-2023

All Science Journal Classification (ASJC) codes

  • Surgery
  • Clinical Neurology

Fingerprint

Dive into the research topics of 'Persistent contrast-filling in the woven endobridge device three months after its implantation for cerebral aneurysm: Incidence, predictive factors, and outcome'. Together they form a unique fingerprint.

Cite this