TY - JOUR
T1 - Stability of radiomic features from positron emission tomography images
T2 - a phantom study comparing advanced reconstruction algorithms and ordered subset expectation maximization
AU - Shiiba, Takuro
AU - Watanabe, Masanori
N1 - Publisher Copyright:
© Australasian College of Physical Scientists and Engineers in Medicine 2024.
PY - 2024/9
Y1 - 2024/9
N2 - In this study, we compared the repeatability and reproducibility of radiomic features obtained from positron emission tomography (PET) images according to the reconstruction algorithm used—advanced reconstruction algorithms, such as HYPER iterative (IT), HYPER deep learning reconstruction (DLR), and HYPER deep progressive reconstruction (DPR), or traditional Ordered Subset Expectation Maximization (OSEM)—to understand the potential variations and implications of using advanced reconstruction techniques in PET-based radiomics. We used a heterogeneous phantom with acrylic spherical beads (4- or 8-mm diameter) filled with 18F. PET images were acquired and reconstructed using OSEM, IT, DLR, and DPR. Original and wavelet radiomic features were calculated using SlicerRadiomics. Radiomic feature repeatability was assessed using the Coefficient of Variance (COV) and intraclass correlation coefficient (ICC), and inter-acquisition time reproducibility was assessed using the concordance correlation coefficient (CCC). For the 4- and 8-mm diameter beads phantom, the proportion of radiomic features with a COV < 10% was equivocal or higher for the advanced reconstruction algorithm than for OSEM. ICC indicated that advanced methods generally outperformed OSEM in repeatability, except for the original features of the 8-mm beads phantom. In the inter-acquisition time reproducibility analysis, the combinations of 3 and 5 min exhibited the highest reproducibility in both phantoms, with IT and DPR showing the highest proportion of radiomic features with CCC > 0.8. Advanced reconstruction methods provided enhanced stability of radiomic features compared with OSEM, suggesting their potential for optimal image reconstruction in PET-based radiomics, offering potential benefits in clinical diagnostics and prognostics.
AB - In this study, we compared the repeatability and reproducibility of radiomic features obtained from positron emission tomography (PET) images according to the reconstruction algorithm used—advanced reconstruction algorithms, such as HYPER iterative (IT), HYPER deep learning reconstruction (DLR), and HYPER deep progressive reconstruction (DPR), or traditional Ordered Subset Expectation Maximization (OSEM)—to understand the potential variations and implications of using advanced reconstruction techniques in PET-based radiomics. We used a heterogeneous phantom with acrylic spherical beads (4- or 8-mm diameter) filled with 18F. PET images were acquired and reconstructed using OSEM, IT, DLR, and DPR. Original and wavelet radiomic features were calculated using SlicerRadiomics. Radiomic feature repeatability was assessed using the Coefficient of Variance (COV) and intraclass correlation coefficient (ICC), and inter-acquisition time reproducibility was assessed using the concordance correlation coefficient (CCC). For the 4- and 8-mm diameter beads phantom, the proportion of radiomic features with a COV < 10% was equivocal or higher for the advanced reconstruction algorithm than for OSEM. ICC indicated that advanced methods generally outperformed OSEM in repeatability, except for the original features of the 8-mm beads phantom. In the inter-acquisition time reproducibility analysis, the combinations of 3 and 5 min exhibited the highest reproducibility in both phantoms, with IT and DPR showing the highest proportion of radiomic features with CCC > 0.8. Advanced reconstruction methods provided enhanced stability of radiomic features compared with OSEM, suggesting their potential for optimal image reconstruction in PET-based radiomics, offering potential benefits in clinical diagnostics and prognostics.
KW - Ordered subset expectation maximization
KW - Positron emission tomography
KW - Radiomic features
KW - Reconstruction algorithms
KW - Stability
UR - http://www.scopus.com/inward/record.url?scp=85190456796&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85190456796&partnerID=8YFLogxK
U2 - 10.1007/s13246-024-01416-x
DO - 10.1007/s13246-024-01416-x
M3 - Article
C2 - 38625624
AN - SCOPUS:85190456796
SN - 2662-4729
VL - 47
SP - 929
EP - 937
JO - Physical and Engineering Sciences in Medicine
JF - Physical and Engineering Sciences in Medicine
IS - 3
ER -