Swallowing computed tomography: Dose estimation in a phantom study conducted at various patient reclining angles

Masanao Kobayashi, Yasuki Asada, Kosuke Matsubara, Kazuyuki Minami, Shouichi Suzuki, Yuta Matsunaga, Tomonobu Haba, Ai Kawaguchi, Tomihiko Daioku, Hiroshi Toyama, Ryoichi Kato

Research output: Contribution to journalArticle

Abstract

Swallowing computed tomography (SCT) is a relatively new technique for the morphological and kinematic analyses of swallowing. However, no optimal scan protocols are available till date. We conducted the present SCT study to estimate the patient dose at various patient reclining positions. A RANDO phantom with a thermoluminescent dosemeter was placed on a hard Table board in a semi-reclining position at the centre and off-centre. According to predetermined scan protocols, irradiation was performed to acquire scanograms at reclining angles of 55° and 65°. The effective dose was the lowest at the centre 45° (3.8 mSv) reclining angle. Comparison between the off-centre (4.6 mSv at 55°, 6.8 mSv at 65°) and centre (4.5 mSv, 5.8 mSv) values suggested that the off-centre position is undesirable with regard to the patient dose. Accordingly, we believe that SCT methods must be revised on the basis of these factors.

Original languageEnglish
Article numberncx078
Pages (from-to)87-94
Number of pages8
JournalRadiation protection dosimetry
Volume178
Issue number1
DOIs
Publication statusPublished - 01-01-2018

Fingerprint

Deglutition
tomography
Tomography
dosage
Biomechanical Phenomena
kinematics
irradiation
estimates

All Science Journal Classification (ASJC) codes

  • Radiological and Ultrasound Technology
  • Radiation
  • Radiology Nuclear Medicine and imaging
  • Public Health, Environmental and Occupational Health

Cite this

Kobayashi, Masanao ; Asada, Yasuki ; Matsubara, Kosuke ; Minami, Kazuyuki ; Suzuki, Shouichi ; Matsunaga, Yuta ; Haba, Tomonobu ; Kawaguchi, Ai ; Daioku, Tomihiko ; Toyama, Hiroshi ; Kato, Ryoichi. / Swallowing computed tomography : Dose estimation in a phantom study conducted at various patient reclining angles. In: Radiation protection dosimetry. 2018 ; Vol. 178, No. 1. pp. 87-94.
@article{94a367a1775d454c8ca5de543b3098fe,
title = "Swallowing computed tomography: Dose estimation in a phantom study conducted at various patient reclining angles",
abstract = "Swallowing computed tomography (SCT) is a relatively new technique for the morphological and kinematic analyses of swallowing. However, no optimal scan protocols are available till date. We conducted the present SCT study to estimate the patient dose at various patient reclining positions. A RANDO phantom with a thermoluminescent dosemeter was placed on a hard Table board in a semi-reclining position at the centre and off-centre. According to predetermined scan protocols, irradiation was performed to acquire scanograms at reclining angles of 55° and 65°. The effective dose was the lowest at the centre 45° (3.8 mSv) reclining angle. Comparison between the off-centre (4.6 mSv at 55°, 6.8 mSv at 65°) and centre (4.5 mSv, 5.8 mSv) values suggested that the off-centre position is undesirable with regard to the patient dose. Accordingly, we believe that SCT methods must be revised on the basis of these factors.",
author = "Masanao Kobayashi and Yasuki Asada and Kosuke Matsubara and Kazuyuki Minami and Shouichi Suzuki and Yuta Matsunaga and Tomonobu Haba and Ai Kawaguchi and Tomihiko Daioku and Hiroshi Toyama and Ryoichi Kato",
year = "2018",
month = "1",
day = "1",
doi = "10.1093/rpd/ncx078",
language = "English",
volume = "178",
pages = "87--94",
journal = "Radiation Protection Dosimetry",
issn = "0144-8420",
publisher = "Oxford University Press",
number = "1",

}

Kobayashi, M, Asada, Y, Matsubara, K, Minami, K, Suzuki, S, Matsunaga, Y, Haba, T, Kawaguchi, A, Daioku, T, Toyama, H & Kato, R 2018, 'Swallowing computed tomography: Dose estimation in a phantom study conducted at various patient reclining angles', Radiation protection dosimetry, vol. 178, no. 1, ncx078, pp. 87-94. https://doi.org/10.1093/rpd/ncx078

Swallowing computed tomography : Dose estimation in a phantom study conducted at various patient reclining angles. / Kobayashi, Masanao; Asada, Yasuki; Matsubara, Kosuke; Minami, Kazuyuki; Suzuki, Shouichi; Matsunaga, Yuta; Haba, Tomonobu; Kawaguchi, Ai; Daioku, Tomihiko; Toyama, Hiroshi; Kato, Ryoichi.

In: Radiation protection dosimetry, Vol. 178, No. 1, ncx078, 01.01.2018, p. 87-94.

Research output: Contribution to journalArticle

TY - JOUR

T1 - Swallowing computed tomography

T2 - Dose estimation in a phantom study conducted at various patient reclining angles

AU - Kobayashi, Masanao

AU - Asada, Yasuki

AU - Matsubara, Kosuke

AU - Minami, Kazuyuki

AU - Suzuki, Shouichi

AU - Matsunaga, Yuta

AU - Haba, Tomonobu

AU - Kawaguchi, Ai

AU - Daioku, Tomihiko

AU - Toyama, Hiroshi

AU - Kato, Ryoichi

PY - 2018/1/1

Y1 - 2018/1/1

N2 - Swallowing computed tomography (SCT) is a relatively new technique for the morphological and kinematic analyses of swallowing. However, no optimal scan protocols are available till date. We conducted the present SCT study to estimate the patient dose at various patient reclining positions. A RANDO phantom with a thermoluminescent dosemeter was placed on a hard Table board in a semi-reclining position at the centre and off-centre. According to predetermined scan protocols, irradiation was performed to acquire scanograms at reclining angles of 55° and 65°. The effective dose was the lowest at the centre 45° (3.8 mSv) reclining angle. Comparison between the off-centre (4.6 mSv at 55°, 6.8 mSv at 65°) and centre (4.5 mSv, 5.8 mSv) values suggested that the off-centre position is undesirable with regard to the patient dose. Accordingly, we believe that SCT methods must be revised on the basis of these factors.

AB - Swallowing computed tomography (SCT) is a relatively new technique for the morphological and kinematic analyses of swallowing. However, no optimal scan protocols are available till date. We conducted the present SCT study to estimate the patient dose at various patient reclining positions. A RANDO phantom with a thermoluminescent dosemeter was placed on a hard Table board in a semi-reclining position at the centre and off-centre. According to predetermined scan protocols, irradiation was performed to acquire scanograms at reclining angles of 55° and 65°. The effective dose was the lowest at the centre 45° (3.8 mSv) reclining angle. Comparison between the off-centre (4.6 mSv at 55°, 6.8 mSv at 65°) and centre (4.5 mSv, 5.8 mSv) values suggested that the off-centre position is undesirable with regard to the patient dose. Accordingly, we believe that SCT methods must be revised on the basis of these factors.

UR - http://www.scopus.com/inward/record.url?scp=85041286097&partnerID=8YFLogxK

UR - http://www.scopus.com/inward/citedby.url?scp=85041286097&partnerID=8YFLogxK

U2 - 10.1093/rpd/ncx078

DO - 10.1093/rpd/ncx078

M3 - Article

C2 - 28633341

AN - SCOPUS:85041286097

VL - 178

SP - 87

EP - 94

JO - Radiation Protection Dosimetry

JF - Radiation Protection Dosimetry

SN - 0144-8420

IS - 1

M1 - ncx078

ER -