TY - JOUR
T1 - The protective effect of tight-fitting powered air-purifying respirators during chest compressions
AU - Goto, Yukari
AU - Jingushi, Naruhiro
AU - Hiraiwa, Hiroaki
AU - Ogawa, Hiroaki
AU - Sakai, Yoshinori
AU - Kasugai, Daisuke
AU - Tanaka, Taku
AU - Higashi, Michiko
AU - Yamamoto, Takanori
AU - Numaguchi, Atsushi
N1 - Publisher Copyright:
© 2021 Elsevier Inc.
PY - 2021/11
Y1 - 2021/11
N2 - Background: Airborne personal protective equipment is required for healthcare workers when performing aerosol-generating procedures on patients with infectious diseases. Chest compressions, one of the main components of cardiopulmonary resuscitation, require intense and dynamic movements of the upper body. We aimed to investigate the protective effect of tight-fitting powered air-purifying respirators (PAPRs) during chest compressions. Methods: This single-center simulation study was performed from February 2021 to March 2021. The simulated workplace protection factor (SWPF) is the concentration ratio of ambient particles and particles inside the PAPR mask; this value indicates the level of protection provided by a respirator when subjected to a simulated work environment. Participants performed continuous chest compressions three times for 2 min each time, with a 4-min break between each session. We measured the SWPF of the tight-fitting PAPR during chest compression in real-time mode. The primary outcome was the ratio of any failure of protection (SWPF <500) during the chest compression sessions. Results: Fifty-four participants completed the simulation. Overall, 78% (n = 42) of the participants failed (the measured SWPF value was less than 500) at least one of the three sessions of chest compressions. The median value and interquartile range of the SWPF was 4304 (685–16,191). There were no reports of slipping down of the respirator or mechanical failure during chest compressions. Conclusions: Although the median SWPF value was high during chest compressions, the tight-fitting PAPR did not provide adequate protection.
AB - Background: Airborne personal protective equipment is required for healthcare workers when performing aerosol-generating procedures on patients with infectious diseases. Chest compressions, one of the main components of cardiopulmonary resuscitation, require intense and dynamic movements of the upper body. We aimed to investigate the protective effect of tight-fitting powered air-purifying respirators (PAPRs) during chest compressions. Methods: This single-center simulation study was performed from February 2021 to March 2021. The simulated workplace protection factor (SWPF) is the concentration ratio of ambient particles and particles inside the PAPR mask; this value indicates the level of protection provided by a respirator when subjected to a simulated work environment. Participants performed continuous chest compressions three times for 2 min each time, with a 4-min break between each session. We measured the SWPF of the tight-fitting PAPR during chest compression in real-time mode. The primary outcome was the ratio of any failure of protection (SWPF <500) during the chest compression sessions. Results: Fifty-four participants completed the simulation. Overall, 78% (n = 42) of the participants failed (the measured SWPF value was less than 500) at least one of the three sessions of chest compressions. The median value and interquartile range of the SWPF was 4304 (685–16,191). There were no reports of slipping down of the respirator or mechanical failure during chest compressions. Conclusions: Although the median SWPF value was high during chest compressions, the tight-fitting PAPR did not provide adequate protection.
UR - http://www.scopus.com/inward/record.url?scp=85107636990&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85107636990&partnerID=8YFLogxK
U2 - 10.1016/j.ajem.2021.06.012
DO - 10.1016/j.ajem.2021.06.012
M3 - Article
C2 - 34118785
AN - SCOPUS:85107636990
SN - 0735-6757
VL - 49
SP - 172
EP - 177
JO - American Journal of Emergency Medicine
JF - American Journal of Emergency Medicine
ER -