Vibration assessment for thrombus formation in the centrifugal pump

Tadashi Nakazawa, Kenzo Makinouchi, Yoshiyuki Takami, Julie Glueck, Eiki Tayama, Yukihiko Nosé

Research output: Contribution to journalArticlepeer-review

4 Citations (Scopus)

Abstract

To clarify the correlation of vibration and thrombus formation inside a rotary blood pump, 40 preliminary vibration studies were performed on pivot bearing centrifugal pumps. No such studies were found in the literature. The primary data acquisition equipment included an accelerometer (Isotron PE accelerometer, ENDEVCO, San Juan Capistrano, CA, U.S.A.), digitizing oscilloscope (TDS 420, Tektronix Inc., Pittsfield, MA, U.S.A.), and pivot bearing centrifugal pumps. The pump impeller was coupled magnetically to the driver magnet. The accelerometer was mounted on the top of the pump casing to sense radial and axial accelerations. To simulate the 3 common areas of thrombus formation, a piece of silicone rubber was attached to each of the following 3 locations as described: a circular shape on the center bottom of the impeller (CI), an eccentric shape on the bottom of the impeller (EI), and a circular shape on the center bottom casing (CC). A fast Fourier transform (FFT) method at 5 L/min against 100 mm Hg, with a pump rotating speed of 1,600 rpm was used. The frequency response of the vibration sensors used spans of 40 Hz to 2 kHz. The frequency domain was already integrated into the oscilloscope, allowing for comparison of the vibration results. The area of frequency domain at a radial direction was 206 ± 12.7 mVHz in CI, 239.5 ± 12.1 mVHz in EI, 365 ± 12.9 mVHz in CC, and 163 ± 7.9 mVHz in the control (control vs. CI p = 0.07, control vs. EI p < 0.001, control vs. CC p < 0.001, EI vs. CC p < 0.001, CI vs. CC p < 0.001). Three types of imitation thrombus formations were roughly distinguishable. These results suggested the possibility of detecting thrombus formation using vibration signals, and these studies revealed the usefulness of vibration monitoring to detect thrombus formation in a centrifugal pump.

Original languageEnglish
Pages (from-to)318-322
Number of pages5
JournalArtificial Organs
Volume21
Issue number4
DOIs
Publication statusPublished - 1997
Externally publishedYes

All Science Journal Classification (ASJC) codes

  • Bioengineering
  • Medicine (miscellaneous)
  • Biomaterials
  • Biomedical Engineering

Fingerprint

Dive into the research topics of 'Vibration assessment for thrombus formation in the centrifugal pump'. Together they form a unique fingerprint.

Cite this