抄録
123I-metaiodobenzylguanidine cardiac scintigraphy (MIBG) is a useful imaging technique for the diagnosis of dementia with Lewy bodies (DLB). However, MIBG has a serious disadvantage in that it demands a long examination time. The objective of this study was to evaluate statistically the usefulness of the heart/mediastinum ratio (H/M) from the early phase of MIBG for the differential diagnosis of DLB. In total, 113 patients were examined, including 32 non-DLB (19 with Alzheimer's dementia) and 79 DLB patients. The mean early-H/M ratio was 2.83 in the non-DLB group and 1.95 in the DLB group. The mean delayed-H/M ratio was 3.0 in the non-DLB group and 1.76 in the DLB group. With a cutoff point of 2.27 on early images, the sensitivity, specificity, and diagnostic accuracy were 65%, 94%, and 73%, respectively, and the area under the curve was 0.82, indicating moderate accuracy. This analysis indicates that images from the early phase of MIBG alone are sufficient for the differential diagnosis of Alzheimer's disease and DLB.
| 本文言語 | 英語 |
|---|---|
| ページ(範囲) | 75-79 |
| ページ数 | 5 |
| ジャーナル | Psychiatry Research - Neuroimaging |
| 巻 | 261 |
| DOI | |
| 出版ステータス | 出版済み - 30-03-2017 |
| 外部発表 | はい |
UN SDG
この成果は、次の持続可能な開発目標に貢献しています
-
SDG 3 すべての人に健康と福祉を
All Science Journal Classification (ASJC) codes
- 神経科学(その他)
- 放射線学、核医学およびイメージング
- 精神医学および精神衛生
フィンガープリント
「123I-metaiodobenzylguanidine myocardial scintigraphy with early images alone is useful for the differential diagnosis of dementia with Lewy bodies」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。引用スタイル
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver