3D automatic exposure control for 64-detector row CT: Radiation dose reduction in chest phantom study

Keiko Matsumoto, Yoshiharu Ohno, Hisanobu Koyama, Atsushi Kono, Hiroyasu Inokawa, Yumiko Onishi, Munenobu Nogami, Daisuke Takenaka, Tsutomu Araki, Kazuro Sugimura

研究成果: Article査読

16 被引用数 (Scopus)

抄録

Purpose: The purpose of this study was to determine the utility of three-dimensional (3D) automatic exposure control (AEC) for low-dose CT examination in a chest phantom study. Materials and methods: A chest CT phantom including simulated focal ground-glass opacities (GGOs) and nodules was scanned with a 64-detector row CT with and without AEC. Performance of 3D AEC included changing targeted standard deviations (SDs) of image noise from scout view. To determine the appropriate targeted SD number for identification, the capability of overall identification with the CT protocol adapted to each of the targeted SDs was compared with that obtained with CT without AEC by means of receiver operating characteristic analysis. Results: When targeted SD values equal to or higher than 250 were used, areas under the curve (Azs) of nodule identification with CT protocol using AEC were significantly smaller than that for CT protocol without AEC (p < 0.05). When targeted SD numbers at equal to or more than 180 were adapted, Azs of CT protocol with AEC had significantly smaller than that without AEC (p < 0.05). Conclusion: This phantom study shows 3D AEC is useful for low-dose lung CT examination, and can reduce the radiation dose while maintaining good identification capability and good image quality.

本文言語English
ページ(範囲)522-527
ページ数6
ジャーナルEuropean journal of radiology
77
3
DOI
出版ステータスPublished - 03-2011
外部発表はい

All Science Journal Classification (ASJC) codes

  • 放射線学、核医学およびイメージング

フィンガープリント

「3D automatic exposure control for 64-detector row CT: Radiation dose reduction in chest phantom study」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル