889 days of support on hydrodynamic bearing rotation mode of the DuraHeart™ for bridge-to-heart transplantation

  • Yorihiko Matsumoto
  • , Tomoyuki Fujita
  • , Satsuki Fukushima
  • , Hiroki Hata
  • , Yusuke Shimahara
  • , Yuta Kume
  • , Kizuku Yamashita
  • , Kensuke Kuroda
  • , Seiko Nakajima
  • , Haruki Sunami
  • , Takuma Sato
  • , Osamu Seguchi
  • , Masanobu Yanase
  • , Norihide Fukushima
  • , Junjiro Kobayashi

研究成果: ジャーナルへの寄稿 › 学術論文 › 査読

抄録

A 49-year-old man with ischemic cardiomyopathy and tricuspid regurgitation underwent a DuraHeart implantation and tricuspid annuloplasty for bridge-to-heart transplantation. On postoperative day 393, the magnetic levitation system suddenly broke down, and the pump system went into hydrodynamic bearing rotation (HD) mode without causing relevant symptoms. The controller was exchanged with one that adapted to the HD mode. No significant hemodynamic changes or indications of hemolysis were observed. On postoperative day 982, the pump temporarily stopped nine times. The patient refused pump exchange despite our strong recommendation for it. After 1283 days of DuraHeart support (889 days in HD mode) without hemolysis or neurologic events, he underwent heart transplantation. The DuraHeart manufacturer’s analysis revealed much damage to the insulation and fatigue fractures of the conductors, which had resulted in temporary cessation of function and failure of the magnetic levitation system. This was a rare case of long-term support under the DuraHeart HD mode.

本文言語英語
ページ(範囲)270-273
ページ数4
ジャーナルJournal of Artificial Organs
巻20
号3
DOI
出版ステータス出版済み - 01-09-2017
外部発表はい

All Science Journal Classification (ASJC) codes

  • 医学(その他)
  • 生体材料
  • 生体医工学
  • 循環器および心血管医学

フィンガープリント

「889 days of support on hydrodynamic bearing rotation mode of the DuraHeart™ for bridge-to-heart transplantation」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル