TY - JOUR
T1 - A 65-nm CMOS Fully Integrated Analysis Platform Using an On-Chip Vector Network Analyzer and a Transmission-Line-Based Detection Window for Analyzing Circulating Tumor Cell and Exosome
AU - Niitsu, Kiichi
AU - Nakanishi, Taiki
AU - Murakami, Shunya
AU - Matsunaga, Maya
AU - Kobayashi, Atsuki
AU - Karim, Nissar Mohammad
AU - Ito, Jun
AU - Ozawa, Naoya
AU - Hase, Tetsunari
AU - Tanaka, Hiromasa
AU - Sato, Mitsuo
AU - Kondo, Hiroki
AU - Ishikawa, Kenji
AU - Odaka, Hidefumi
AU - Hasegawa, Yoshinori
AU - Hori, Masaru
AU - Nakazato, Kazuo
N1 - Publisher Copyright:
© 2007-2012 IEEE.
PY - 2019/4
Y1 - 2019/4
N2 - A fully integrated CMOS circuit based on a vector network analyzer and a transmission-line-based detection window for circulating tumor cell (CTC) and exosome analysis is presented for the first time. We have introduced a fully integrated architecture, which eliminates the undesired parasitic components and enables high-sensitivity, to analyze extremely low-concentration CTC in blood. The detection window was designed on the high-sensitive coplanar waveguide line. To validate the operation of the proposed system, a test chip was fabricated using 65-nm CMOS technology. Measurements were performed after adding a tiny lump of silicone or a droplet of water on its detection window. The measured results show S-21 degradation of-1.96 dB and-6.04 dB for the silicone and the droplet, respectively, at 1.4 GHz. In addition, in another measurement using magnetic beads, it is confirmed that the proposed circuit can analyze even low concentrations of 20 beads/μL. As well as microbeads, measurement with CTCs was successfully demonstrated.
AB - A fully integrated CMOS circuit based on a vector network analyzer and a transmission-line-based detection window for circulating tumor cell (CTC) and exosome analysis is presented for the first time. We have introduced a fully integrated architecture, which eliminates the undesired parasitic components and enables high-sensitivity, to analyze extremely low-concentration CTC in blood. The detection window was designed on the high-sensitive coplanar waveguide line. To validate the operation of the proposed system, a test chip was fabricated using 65-nm CMOS technology. Measurements were performed after adding a tiny lump of silicone or a droplet of water on its detection window. The measured results show S-21 degradation of-1.96 dB and-6.04 dB for the silicone and the droplet, respectively, at 1.4 GHz. In addition, in another measurement using magnetic beads, it is confirmed that the proposed circuit can analyze even low concentrations of 20 beads/μL. As well as microbeads, measurement with CTCs was successfully demonstrated.
KW - Circulating tumor cell (CTC)
KW - CMOS
KW - coplanar waveguide (CPW) line
KW - dielectric properties
KW - exosome
KW - on-chip VNA
UR - https://www.scopus.com/pages/publications/85057161020
UR - https://www.scopus.com/pages/publications/85057161020#tab=citedBy
U2 - 10.1109/TBCAS.2018.2882472
DO - 10.1109/TBCAS.2018.2882472
M3 - Article
C2 - 30475728
AN - SCOPUS:85057161020
SN - 1932-4545
VL - 13
SP - 470
EP - 479
JO - IEEE Transactions on Biomedical Circuits and Systems
JF - IEEE Transactions on Biomedical Circuits and Systems
IS - 2
M1 - 8540882
ER -