A case of IVUS catheter detachment after becoming entrapped with stent struts implanted two years before

Hideki Kawai, Miho Tanaka, Yoshihiro Murata, Masayoshi Koyasu, Hideki Horibe, Masanobu Yanase, Kenji Takemoto, Makoto Nonokawa, Seiji Shimizu, Masato Watarai

研究成果: Article査読

2 被引用数 (Scopus)

抄録

A 71-year-old man who had undergone PCI several times previously was admitted to our hospital to treat a distal lesion of the RCA. At first we crossed a PT Graphix® into the culprit lesion smoothly. Next we tried to cross an IVUS catheter, but could only pass it into the middle of the RCA. It became entrapped and subsequently detached when we tried to pull it out from the RCA. In spite of using a goose neck snare, we couldn't remove the ruptured catheter. An emergent operation revealed that the IVUS catheter was twined around the struts of the stent which had been implanted in the middle of the RCA two years before.

本文言語English
ページ(範囲)333-337
ページ数5
ジャーナルJapanese Journal of Interventional Cardiology
21
4
出版ステータスPublished - 2006
外部発表はい

All Science Journal Classification (ASJC) codes

  • Radiology Nuclear Medicine and imaging
  • Cardiology and Cardiovascular Medicine

フィンガープリント 「A case of IVUS catheter detachment after becoming entrapped with stent struts implanted two years before」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。