Aortic dissection false lumen treatment with novel shape memory polymer devices

研究成果: ジャーナルへの寄稿学術論文査読

抄録

In this case series, innovative shape memory polymer (SMP) devices were used to treat the aortic dissection false lumen (FL) of five patients. The open-cell, porous structure of expanded SMP contributes to its radiolucency, low radial force, and conformability. The devices were designed to address the unique needs of treating the FL: sufficiently large volume, detachable and stable to minimize migration risk, segmented for flexibility, and with a range of dimensions. All five cases were technically successful. A decrease in total aortic diameter, true lumen restoration, and FL reduction were observed through 6- to 12-month follow-up. The mean changes in maximum aorta diameter from preoperative measurements were −5.8 ± 6.9 mm (30 days, n = 5), −12.0 ± 9.4 mm (6 months, n = 4), and −9.3 ± 0.8 mm (12 months, n = 2). No SMP device/procedure-related complications occurred.

本文言語英語