To obtain objective and concrete data by physically assessing the quality of breast magnetic resonance images based on the fat-suppression effect by the modified Dixon method (mDixon) and frequency-selective fat suppression (e-Thrive) using an original lipid-content breast phantom that could easily reveal the influence of non-uniform fat suppression in breast magnetic resonance imaging. The fat-suppression uniformity was approximately seven times superior when using mDixon compared with when using e-Thrive. mDixon appears to have a significant advantage.
| 本文言語 | 英語 |
|---|---|
| ページ(範囲) | 350-355 |
| ページ数 | 6 |
| ジャーナル | Magnetic Resonance in Medical Sciences |
| 巻 | 17 |
| 号 | 4 |
| DOI |
|
| 出版ステータス | 出版済み - 2018 |
| 外部発表 | はい |