T1 - Can magnetic resonance imaging after cranioplasty using titanium mesh detect brain tumors?
AU - Takatsu, Yasuo
AU - Yoshida, Rei
AU - Yamatani, Yuya
AU - Kanno, Mikihisa
AU - Miyati, Tosiaki
N1 - Funding Information:
The titanium mesh was provided by Bear Medic Corporation (Ibaraki, Japan).
Publisher Copyright:
© 2022, Australasian College of Physical Scientists and Engineers in Medicine.
PY - 2023/3
Y1 - 2023/3
N2 - This study determined the dependence of the concentration and position of contrast-enhanced tumors on the radio frequency (RF)-shielding effect of titanium mesh using the contrast-to-noise ratio (CNR) in magnetic resonance imaging (MRI). A phantom was constructed by filling a plastic container with manganese chloride tetrahydrate and agar. Four cellophane cylindrical containers were arranged from the end of the plastic container, and the brain tumor model was filled with gadobutrol diluted with NaCl, with molarity values of 0.2–1.0 mmol/L. The titanium mesh board was set on the left side of the phantom. Images were acquired using a 1.5-T MRI as well as two-dimensional spin-echo (2D SE) and three-dimensional fast spoiled gradient echo (3D FSPGR) sequences. CNR was calculated using the signal intensity values of the tumor model, surrounding area of the brain model, and background noise. Furthermore, the fractional change in CNR was calculated using values of CNR with and without the mesh. Moreover, a profile of CNR was created. The fractional change in CNR decreased at the brain tumor positions present near the mesh and at a contrast medium concentration of approximately ≤ 0.5 mmol/L in 2D SE and ≤ 0.25 mmol/L in 3D FSPGR. According to the CNR profiles, directly under the mesh, almost all contrast concentrations in 2D SE was unrecognizable; however, at a concentration of ≥ 0.5 mmol/L in 3D FSPGR was recognizable.
AB - This study determined the dependence of the concentration and position of contrast-enhanced tumors on the radio frequency (RF)-shielding effect of titanium mesh using the contrast-to-noise ratio (CNR) in magnetic resonance imaging (MRI). A phantom was constructed by filling a plastic container with manganese chloride tetrahydrate and agar. Four cellophane cylindrical containers were arranged from the end of the plastic container, and the brain tumor model was filled with gadobutrol diluted with NaCl, with molarity values of 0.2–1.0 mmol/L. The titanium mesh board was set on the left side of the phantom. Images were acquired using a 1.5-T MRI as well as two-dimensional spin-echo (2D SE) and three-dimensional fast spoiled gradient echo (3D FSPGR) sequences. CNR was calculated using the signal intensity values of the tumor model, surrounding area of the brain model, and background noise. Furthermore, the fractional change in CNR was calculated using values of CNR with and without the mesh. Moreover, a profile of CNR was created. The fractional change in CNR decreased at the brain tumor positions present near the mesh and at a contrast medium concentration of approximately ≤ 0.5 mmol/L in 2D SE and ≤ 0.25 mmol/L in 3D FSPGR. According to the CNR profiles, directly under the mesh, almost all contrast concentrations in 2D SE was unrecognizable; however, at a concentration of ≥ 0.5 mmol/L in 3D FSPGR was recognizable.
UR - http://www.scopus.com/inward/record.url?scp=85143405640&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85143405640&partnerID=8YFLogxK
U2 - 10.1007/s13246-022-01200-9
DO - 10.1007/s13246-022-01200-9
M3 - Article
AN - SCOPUS:85143405640
SN - 2662-4729
VL - 46
SP - 109
EP - 118
JO - Physical and Engineering Sciences in Medicine
JF - Physical and Engineering Sciences in Medicine
IS - 1
ER -