Comparative evaluation of absorbed doses in silver wedge-filter-based chest CT versus chest radiography: A Monte Carlo study

研究成果: ジャーナルへの寄稿学術論文査読

抄録

In the silver (Ag) wedge-filter-fitted computed tomography (CT) scanner, we estimated organ doses during chest CT examinations and chest radiography to assess the Ag-filter-based dose reduction in six scan protocols for chest CT: three conventional protocols with standard filters (clinical standard, low-dose, and ultra-low-dose) and three corresponding Ag-filter-based protocols; the tube current was adjusted to maintain similar standard deviations (SD) and two protocols for chest radiography: posteroanterior (PA) projection (medical checkups) and posteroanterior with right-left (PA + RL) projection (clinical protocols). Using the Monte Carlo simulation software (ImpactMC), we estimated absorbed doses by simulating photon transport in computational phantoms (adult males and females) based on average values for organ doses (lungs, breast tissue, heart, stomach, and liver) and the maximum value for the skin dose. Despite similar SD, the organ and tissue doses in the Ag filter-based ultra-low-dose chest CT protocol (vs. no-Ag filter) were lower for females: skin (39.6 %), lung (17.4 %), breast tissue (20.1 %), heart (16.5 %), stomach (14.4 %), and liver (16.7 %), with similar dose reductions across other scan protocols. In the ultra-low-dose Ag-filter-based chest CT protocol, the absorbed doses resembled those of PA + RL radiography (clinical protocol). During chest CT examinations, the Ag filter-based effective reduction of the radiation dose conferred a potentially significant impact on the risk–benefit balance in CT-based lung cancer screening; this potentially influences the screening targets and overall effectiveness. By balancing radiation risk and diagnostic benefits, the Ag filter facilitates a paradigm shift from radiography to ultra-low-dose CT examinations in conventional diagnosis and follow-up.

本文言語英語
論文番号113274
ジャーナルRadiation Physics and Chemistry
239
DOI
出版ステータス出版済み - 02-2026
外部発表はい

UN SDG

この成果は、次の持続可能な開発目標に貢献しています

  1. SDG 3 - すべての人に健康と福祉を
    SDG 3 すべての人に健康と福祉を

All Science Journal Classification (ASJC) codes

  • 放射線

フィンガープリント

「Comparative evaluation of absorbed doses in silver wedge-filter-based chest CT versus chest radiography: A Monte Carlo study」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル