TY - JOUR
T1 - Comparative evaluation of zoomed and parallel diffusion-weighted imaging
T2 - A phantom study
AU - Miyati, Tosiaki
AU - Iwase, Akiyoshi
AU - Takatsu, Yasuo
AU - Ohno, Naoki
AU - Kitao, Azusa
N1 - Publisher Copyright:
© 2026 Published by Scientific Scholar on behalf of Journal of Clinical Imaging Science. This is an open-access article distributed under the terms of the Creative Commons Attribution-Non Commercial-Share Alike 4.0 License, which allows others to remix, transform, and build upon the work non-commercially, as long as the author is credited and the new creations are licensed under the identical terms.
PY - 2026
Y1 - 2026
N2 - Objectives: Single-shot echo-planar imaging sequences with parallel imaging are widely used for diffusion-weighted imaging (DWI). However, when small fields of view and higher phase-encoding reduction ratios (PRs) are applied, image quality may deteriorate because of imperfections in the unfolding process of parallel imaging. Previous studies have suggested that Zoomed DWI (Zoom-DWI) may address this limitation. Therefore, we compared Zoom-DWI with parallel DWI (Parallel-DWI) across different PRs to evaluate image quality. Material and Methods: Zoom-DWI and Parallel-DWI were performed using identical acquisition times and imaging parameters at different PRs. Signal-to-noise ratio (SNR), image uniformity, slice thickness, spatial resolution, and image distortion were evaluated. Results: At a PR of 66.6%, the SNR of Zoom-DWI was approximately 1.4–3.3-fold higher than that of Parallel-DWI at the image center, where reconstruction errors in parallel imaging become more pronounced as PR increases. Image uniformity was also higher in Zoom-DWI than in Parallel-DWI, and the difference increased by approximately 41.0% at a PR of 66.6%. Slice thickness, spatial resolution, and image distortion were comparable between the two methods. Conclusion: In this phantom study, Zoom-DWI demonstrated higher SNR and greater image uniformity than Parallel-DWI, particularly at higher PRs and in the central image region.
AB - Objectives: Single-shot echo-planar imaging sequences with parallel imaging are widely used for diffusion-weighted imaging (DWI). However, when small fields of view and higher phase-encoding reduction ratios (PRs) are applied, image quality may deteriorate because of imperfections in the unfolding process of parallel imaging. Previous studies have suggested that Zoomed DWI (Zoom-DWI) may address this limitation. Therefore, we compared Zoom-DWI with parallel DWI (Parallel-DWI) across different PRs to evaluate image quality. Material and Methods: Zoom-DWI and Parallel-DWI were performed using identical acquisition times and imaging parameters at different PRs. Signal-to-noise ratio (SNR), image uniformity, slice thickness, spatial resolution, and image distortion were evaluated. Results: At a PR of 66.6%, the SNR of Zoom-DWI was approximately 1.4–3.3-fold higher than that of Parallel-DWI at the image center, where reconstruction errors in parallel imaging become more pronounced as PR increases. Image uniformity was also higher in Zoom-DWI than in Parallel-DWI, and the difference increased by approximately 41.0% at a PR of 66.6%. Slice thickness, spatial resolution, and image distortion were comparable between the two methods. Conclusion: In this phantom study, Zoom-DWI demonstrated higher SNR and greater image uniformity than Parallel-DWI, particularly at higher PRs and in the central image region.
KW - Image uniformity
KW - Parallel imaging
KW - Phantom study
KW - Signal-to-noise ratio
KW - Zoomed diffusion-weighted imaging
UR - https://www.scopus.com/pages/publications/105046969450
UR - https://www.scopus.com/pages/publications/105046969450#tab=citedBy
U2 - 10.25259/JCIS_92_2025
DO - 10.25259/JCIS_92_2025
M3 - Article
AN - SCOPUS:105046969450
SN - 2156-7514
VL - 16
JO - Journal of Clinical Imaging Science