A new tongue pressure device consisting of a simple and safe disposable probe and manometer has been developed. This report describes a study that examined the validity of the new device, comparing it to a widely used tongue pressure manometer, the Iowa Oral Performance Instrument (IOPI), and to the stable adhered three air-filled bulbs manometry system. The first test compared maximum tongue pressure measured with the new device and the IOPI (13 male, 9 female, 25.0 years). The second test compared maximum tongue pressure and swallowing tongue pressure measured with the new device and the threebulb device (13 male, 9 female, 31.0 years). Significant correlations of maximum tongue pressure were found between the new device and the IOPI in the first test (p <0.05). In the second test, significant correlations of maximum tongue pressure were found between the new device at the anterior and middle sensors (p <0.05) but not at the posterior sensor of the three-bulb device. Significant correlations of swallowing tongue pressure between the new device and the three-bulb device were found (p <0.05). These findings demonstrate that the measurements by the new simple tongue pressure device are closely equivalent to those of the IOPI and three-bulb devices, demonstrating that the new device is capable of accurately measuring the pressure generated by the whole tongue.
| 本文言語 | 英語 |
|---|---|
| ページ(範囲) | 232-237 |
| ページ数 | 6 |
| ジャーナル | Dysphagia |
| 巻 | 26 |
| 号 | 3 |
| DOI | |
| 出版ステータス | 出版済み - 09-2011 |
| 外部発表 | はい |
All Science Journal Classification (ASJC) codes
- 耳鼻咽喉科学
- 消化器病学
- 言語聴覚療法
フィンガープリント
「Comparison of three types of tongue pressure measurement devices」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。引用スタイル
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver