Determination of 3-hydroxyisovalerylcarnitine and other acylcarnitine levels using liquid chromatography-tandem mass spectrometry in serum and urine of a patient with multiple carboxylase deficiency

Yasuhiro Maeda, Tetsuya Ito, Hironori Ohmi, Kyoko Yokoi, Yoko Nakajima, Akihito Ueta, Yukihisa Kurono, Hajime Togari, Naruji Sugiyama

研究成果: Article査読

28 被引用数 (Scopus)

抄録

Due to its increased concentration in blood, 3-hydroxyisovalerylcarnitine (C5OH-I) is an important indicator for the diagnosis of organic acidemias in newborns. However, C5OH-I has not been used as a standard in tandem mass spectrometric (MS/MS) assays because its isolation is difficult. We developed a new synthesis of C5OH-I and investigated its behavior by MS/MS. A method using the multiple reaction monitoring (MRM) mode of MS/MS with HPLC was developed which provides high accuracy, precision and reproducibility. Acylcarnitine profiles in the serum and urine of a patient with multiple carboxylase deficiency (MCD) showed increased levels compared to a healthy patient.

本文言語English
ページ(範囲)154-159
ページ数6
ジャーナルJournal of Chromatography B: Analytical Technologies in the Biomedical and Life Sciences
870
2
DOI
出版ステータスPublished - 15-07-2008
外部発表はい

All Science Journal Classification (ASJC) codes

  • 分析化学
  • 生化学
  • 臨床生化学
  • 細胞生物学

フィンガープリント

「Determination of 3-hydroxyisovalerylcarnitine and other acylcarnitine levels using liquid chromatography-tandem mass spectrometry in serum and urine of a patient with multiple carboxylase deficiency」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル