Development of a new coordinate calibration phantom for a light-section-based optical surface monitoring system

  • Tatsunori Saito
  • , Naoki Hayashi
  • , Hiroshi Amma
  • , Kazuki Onishi
  • , Yuta Muraki
  • , Masashi Nozue

研究成果: ジャーナルへの寄稿学術論文査読

抄録

A calibration phantom made of Derlin requires manual translational and rotational adjustments when calibrating a light-section-based optical surface monitoring system (VOXELAN) with a phantom material that insufficiently reflects the red-slit laser of the system. This study aimed to develop a new calibration phantom using different materials and to propose a procedure that minimizes setup errors. The new phantom, primarily made of PET100, which exhibits good reflectivity without scattering or attenuating the red-slit laser at the phantom surface, was shaped in a manner similar to that of previous designs. The detection accuracy and stability were evaluated using six different regions of interest (ROIs) and compared with previous phantom designs. The coordinate coincidence between the machine and VOXELAN was compared for both phantom designs. The detection accuracy and stability of the new phantom in the reference ROI setting were found to be better than those of previous phantoms. In the lateral, longitudinal, and vertical directions, the coordinate coincidences in translational directions for the previous phantom were obtained at 1.07 ± 0.66, 1.46 ± 0.47, and 0.26 ± 0.83 mm, whereas those for the new phantom were obtained at 0.28 ± 0.21, 0.18 ± 0.30, and − 0.30 ± 0.29 mm, respectively. The rotational errors of the two phantoms were identical. The new phantom exhibited improved detection stability because of its good reflectivity. Additionally, the new placement procedure was linked to the six-degrees-of-freedom couch. A combination of the new phantom and its new placement procedure is suitable for coordinate calibration of VOXELAN.

本文言語英語
ページ(範囲)366-372
ページ数7
ジャーナルRadiological Physics and Technology
16
3
DOI
出版ステータス出版済み - 09-2023
外部発表はい

All Science Journal Classification (ASJC) codes

  • 放射線
  • 理学療法、スポーツ療法とリハビリテーション
  • 放射線学、核医学およびイメージング

フィンガープリント

「Development of a new coordinate calibration phantom for a light-section-based optical surface monitoring system」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル