Development of calibration phantoms for generating quantitative perfusion images using 2D angiography data

Takuya Sakaguchi, Takahiro Natsume, Yuki Kanamori, Takashi Ichihara

研究成果: Conference contribution

1 被引用数 (Scopus)

抄録

Blood flow quantification by perfusion imaging or fractional flow reserve estimation using 2D x-ray angiography data requires scatter correction and beam hardening correction to ensure a linear relationship between the concentration of contrast agent in the human body and the x-ray image signal intensity. These types of correction usually require complicated calibration and numerous correction tables. To simplify correction, we developed a set of spherical epoxy resin phantoms containing different concentrations of barium sulfate. Prior to use, the concentration of iodinated contrast agent that results in the same x-ray absorption as each spherical phantom was determined. These values are hereinafter referred to as the equivalent concentrations of the spherical phantoms. To confirm that the spherical phantoms were suitable for estimating the contrast agent concentration, the curve of the image signal intensity vs. equivalent concentration was compared with the corresponding curve for the contrast agent. The mean difference between the two curves was 18 mgI/mL for 6 sets of experimental conditions. The mean of the p-values obtained in the paired t-tests was greater than 0.05. To validate the accuracy of our method, the spherical phantoms and test samples were placed on the surface of a chest phantom and an x-ray image including all the spherical phantoms and test samples was acquired. Using the relationship between the equivalent concentrations of the spherical phantoms and the image signal intensities, the concentrations of the test samples were estimated from their measured image signal intensities. The estimated and true concentrations of the test samples were then compared. Their cross correlation was R>0.99 for each of the 8 sets of experimental conditions used. The mean of the p-values obtained in the paired t-tests was greater than 0.05. Based on the promising results of these initial tests, we plan to proceed to clinical studies on these phantoms.

本文言語English
ホスト出版物のタイトル2014 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2014
出版社Institute of Electrical and Electronics Engineers Inc.
ISBN(電子版)9781479960972
DOI
出版ステータスPublished - 10-03-2016
イベントIEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2014 - Seattle, United States
継続期間: 08-11-201415-11-2014

出版物シリーズ

名前2014 IEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2014

Conference

ConferenceIEEE Nuclear Science Symposium and Medical Imaging Conference, NSS/MIC 2014
CountryUnited States
CitySeattle
Period08-11-1415-11-14

All Science Journal Classification (ASJC) codes

  • Nuclear and High Energy Physics
  • Radiology Nuclear Medicine and imaging

フィンガープリント 「Development of calibration phantoms for generating quantitative perfusion images using 2D angiography data」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル