Development of immuno-wall devices and a mobile fluorescence reader for on-site sample-to-answer immunoassay

  • Toshihiro Kasama
  • , Yutaka Hasegawa
  • , Haruyasu Kondo
  • , Tsutomu Ozawa
  • , Noritada Kaji
  • , Manabu Tokeshi
  • , Yoshinobu Baba

研究成果: 書籍/レポート タイプへの寄稿 › 会議への寄与

2   !!Link opens in a new tab 被引用数 (Scopus)

抄録

We have developed a novel immunoassay device which has a long and thin wall structure inside a microchannel, hence the name 'immuno-wall device'. Unreacted antigens and antibodies were completely removed by just immersing the device in a washing buffer for 1 minute. The long structure also allowed us to analyze fluorescence intensity by using inexpensive desktop fluorescence reader ($6,000) instead of expensive fluorescence microscopes. By using the immuno-wall devices and the fluorescence reader, the high-sensitivity C-reactive protein assays were performed with sample-in-answer-out in 15 minutes.

本文言語英語
ホスト出版物のタイトル18th International Conference on Miniaturized Systems for Chemistry and Life Sciences, MicroTAS 2014
出版社Chemical and Biological Microsystems Society
ページ935-937
ページ数3
ISBN(電子版)9780979806476
出版ステータス出版済み - 2014
外部発表はい
イベント18th International Conference on Miniaturized Systems for Chemistry and Life Sciences, MicroTAS 2014 - San Antonio, 米国
継続期間: 26-10-2014 → 30-10-2014

出版物シリーズ

名前18th International Conference on Miniaturized Systems for Chemistry and Life Sciences, MicroTAS 2014

会議

会議18th International Conference on Miniaturized Systems for Chemistry and Life Sciences, MicroTAS 2014
国/地域米国
CitySan Antonio
Period26-10-14 → 30-10-14

All Science Journal Classification (ASJC) codes

  • 制御およびシステム工学

フィンガープリント

「Development of immuno-wall devices and a mobile fluorescence reader for on-site sample-to-answer immunoassay」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル