Direct Visualization of External Outflow Graft Obstruction in Durable Left Ventricular Assist Device Using Transthoracic Echocardiography

  • Shin Nagai
  • , Shotaro Komeyama
  • , Toru Kondo
  • , Ryota Morimoto
  • , Shingo Kazama
  • , Hiroaki Hiraiwa
  • , Takahiro Okumura
  • , Tomo Yoshizumi
  • , Masato Mutsuga
  • , Toyoaki Murohara

研究成果: ジャーナルへの寄稿学術論文査読

抄録

External outflow graft obstruction (OGO) in durable left ventricular assist devices (LVADs) is a rare but critical complication. Detecting external OGO can be challenging because of its nonspecific symptoms and a wide variety of differential diagnoses. We report the case of a 45 year old man with LVAD admitted with complaints of recurrent syncope and frequent low-flow alarms. Laboratory tests and chest radiography revealed no specific findings. Transthoracic echocardiography via the left subcostal approach revealed an outflow graft compressed by a low-echo area within a bend relief. This finding led to a diagnosis of external OGO, which was consistent with what was observed on contrast-enhanced computed tomography. The patient underwent surgery and the gelatinous substance compressing the outflow graft was successfully removed. This case highlights the utility of transthoracic echocardiography as a noninvasive and effective method for detecting external OGOs, emphasizing its role as a valuable tool for the initial screening of patients with suspected OGO.

本文言語英語
ページ(範囲)e231-e234
ジャーナルASAIO Journal
71
12
DOI
出版ステータス出版済み - 12-2025
外部発表はい

All Science Journal Classification (ASJC) codes

  • 生物理学
  • バイオエンジニアリング
  • 医学一般
  • 生体材料
  • 生体医工学

フィンガープリント

「Direct Visualization of External Outflow Graft Obstruction in Durable Left Ventricular Assist Device Using Transthoracic Echocardiography」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル