抄録
Objective: We report a case of disruption and migration of the distal portion of the delivery wire from an Enterprise Vascular Reconstruction Device (VRD) during coil embolization for an unruptured cerebral aneurysm. Case presentation: A 63-year-old woman, who had undergone neck clipping for an aneurysm of the right internal carotid-posterior communicating artery 12 years previously, suffered a recurrence at the same location. Stent-assisted coil embolization using an Enterprise VRD was chosen for repair because of its wide neck. After insufficient coil embolization with the initial Enterprise VRD, a second Enterprise VRD was deployed in a stent-in-stent manner. Following deployment of the second stent, a distal section of the delivery wire became disrupted and migrated distal to the aneurysm. Craniotomy was performed for direct surgical removal of the migrated wire because endovascular retrieval with a goose neck snare proved too difficult. Conclusion: Disruption of the delivery wire from an Enterprise VRD is extremely rare but possible. Physicians should be aware of this potential problem in patients with a tortuous parent artery and proceed with caution to avoid excessive stress to the device.
| 本文言語 | 英語 |
|---|---|
| ページ(範囲) | 133-140 |
| ページ数 | 8 |
| ジャーナル | Journal of Neuroendovascular Therapy |
| 巻 | 6 |
| 号 | 2 |
| DOI | |
| 出版ステータス | 出版済み - 2012 |
| 外部発表 | はい |
All Science Journal Classification (ASJC) codes
- 臨床神経学
- 循環器および心血管医学
フィンガープリント
「Disruption and migration of an Enterprise VRD delivery wire during coil embolization for unruptured cerebral aneurysm: Case report」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。引用スタイル
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver