Dual-energy computed tomography of the head: A phantom study assessing axial dose distribution, eye lens dose, and image noise level

Kosuke Matsubara, Hiroki Kawashima, Takashi Hamaguchi, Tadanori Takata, Masanao Kobayashi, Katsuhiro Ichikawa, Kichiro Koshida

研究成果: 書籍/レポート タイプへの寄稿会議への寄与

抄録

The aim of this study was to propose a calibration method for small dosimeters to measure absorbed doses during dual- source dual-energy computed tomography (DECT) and to compare the axial dose distribution, eye lens dose, and image noise level between DE and standard, single-energy (SE) head CT angiography. Three DE (100/Sn140 kVp 80/Sn140 kVp, and 140/80 kVp) and one SE (120 kVp) acquisitions were performed using a second-generation dual-source CT device and a female head phantom, with an equivalent volumetric CT dose index. The axial absorbed dose distribution at the orbital level and the absorbed doses for the eye lens were measured using radiophotoluminescent glass dosimeters. CT attenuation numbers were obtained in the DE composite images and the SE images of the phantom at the orbital level. The doses absorbed at the orbital level and in the eye lens were lower and standard deviations for the CT attenuation numbers were slightly higher in the DE acquisitions than those in the SE acquisition. The anterior surface dose was especially higher in the SE acquisition than that in the DE acquisitions. Thus, DE head CT angiography can be performed with a radiation dose lower than that required for a standard SE head CT angiography, with a slight increase in the image noise level. The 100/Sn140 kVp acquisition revealed the most balanced axial dose distribution. In addition, our proposed method was effective for calibrating small dosimeters to measure absorbed doses in DECT.

本文言語英語
ホスト出版物のタイトルMedical Imaging 2016
ホスト出版物のサブタイトルPhysics of Medical Imaging
編集者Despina Kontos, Joseph Y. Lo, Thomas G. Flohr
出版社SPIE
ISBN(電子版)9781510600188
DOI
出版ステータス出版済み - 2016
イベントMedical Imaging 2016: Physics of Medical Imaging - San Diego, 米国
継続期間: 28-02-201602-03-2016

出版物シリーズ

名前Progress in Biomedical Optics and Imaging - Proceedings of SPIE
9783
ISSN(印刷版)1605-7422

その他

その他Medical Imaging 2016: Physics of Medical Imaging
国/地域米国
CitySan Diego
Period28-02-1602-03-16

All Science Journal Classification (ASJC) codes

  • 電子材料、光学材料、および磁性材料
  • 生体材料
  • 原子分子物理学および光学
  • 放射線学、核医学およびイメージング

フィンガープリント

「Dual-energy computed tomography of the head: A phantom study assessing axial dose distribution, eye lens dose, and image noise level」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル