T1 - Effect of the endothelial shear stress patterns on neointimal proliferation following drug-eluting bioresorbable vascular scaffold implantation
T2 - An optical coherence tomography study
AU - Bourantas, Christos V.
AU - Papafaklis, Michail I.
AU - Kotsia, Anna
AU - Farooq, Vasim
AU - Muramatsu, Takashi
AU - Gomez-Lara, Josep
AU - Zhang, Yao Jun
AU - Iqbal, Javaid
AU - Kalatzis, Fanis G.
AU - Naka, Katerina K.
AU - Fotiadis, Dimitrios I.
AU - Dorange, Cecile
AU - Wang, Jin
AU - Rapoza, Richard
AU - Garcia-Garcia, Hector M.
AU - Onuma, Yoshinobu
AU - Michalis, Lampros K.
AU - Serruys, Patrick W.
N1 - Funding Information:
The ABSORB Cohort B study was sponsored and financially supported by Abbott Vascular. Dr. Bourantas is funded by the Hellenic Cardiological Society . Dr. Dorange is an employee of Abbott Vascular. Drs. Wang and Rapoza are employees of and hold stock in Abbott Vascular. All other authors have reported that they have no relationships relevant to the contents of this paper to disclose.
PY - 2014/3
Y1 - 2014/3
N2 - Objectives: This study sought to investigate the effect of endothelial shear stress (ESS) on neointimal formation following an Absorb bioresorbable vascular scaffold (BVS) (Abbott Vascular, Santa Clara, California) implantation. Background: Cumulative evidence, derived from intravascular ultrasound-based studies, has demonstrated a strong association between local ESS patterns and neointimal formation in bare-metal stents, whereas in drug-eluting stents, there are contradictory data about the effect of ESS on the vessel wall healing process. The effect of ESS on neointimal development following a bioresorbable scaffold implantation remains unclear. Methods: Twelve patients with an obstructive lesion in a relatively straight arterial segment, who were treated with an Absorb BVS and had serial optical coherence tomographic examination at baseline and 1-year follow-up, were included in the current analysis. The optical coherence tomographic data acquired at follow-up were used to reconstruct the scaffolded segment. Blood flow simulation was performed on the luminal surface at baseline defined by the Absorb BVS struts, and the computed ESS was related to the neointima thickness measured at 1-year follow-up. Results: At baseline, the scaffolded segments were exposed to a predominantly low ESS environment (61% of the measured ESS was <1 Pa). At follow-up, the mean neointima thickness was 113 ± 45 μm, whereas the percentage scaffold volume obstruction was 13.1 ± 6.6%. A statistically significant inverse correlation was noted between baseline logarithmic transformed ESS and neointima thickness at 1-year follow-up in all studied segments (correlation coefficient range -0.140 to -0.662). Mixed linear regression analysis between baseline logarithmic transformed ESS and neointima thickness at follow-up yielded a slope of -31 μm/ln(Pa) and a y-intercept of 99 μm. Conclusions: The hemodynamic microenvironment appears to regulate neointimal response following an Absorb BVS implantation. These findings underline the role of the ESS patterns on vessel wall healing and should be taken into consideration in the design of bioresorbable devices.
AB - Objectives: This study sought to investigate the effect of endothelial shear stress (ESS) on neointimal formation following an Absorb bioresorbable vascular scaffold (BVS) (Abbott Vascular, Santa Clara, California) implantation. Background: Cumulative evidence, derived from intravascular ultrasound-based studies, has demonstrated a strong association between local ESS patterns and neointimal formation in bare-metal stents, whereas in drug-eluting stents, there are contradictory data about the effect of ESS on the vessel wall healing process. The effect of ESS on neointimal development following a bioresorbable scaffold implantation remains unclear. Methods: Twelve patients with an obstructive lesion in a relatively straight arterial segment, who were treated with an Absorb BVS and had serial optical coherence tomographic examination at baseline and 1-year follow-up, were included in the current analysis. The optical coherence tomographic data acquired at follow-up were used to reconstruct the scaffolded segment. Blood flow simulation was performed on the luminal surface at baseline defined by the Absorb BVS struts, and the computed ESS was related to the neointima thickness measured at 1-year follow-up. Results: At baseline, the scaffolded segments were exposed to a predominantly low ESS environment (61% of the measured ESS was <1 Pa). At follow-up, the mean neointima thickness was 113 ± 45 μm, whereas the percentage scaffold volume obstruction was 13.1 ± 6.6%. A statistically significant inverse correlation was noted between baseline logarithmic transformed ESS and neointima thickness at 1-year follow-up in all studied segments (correlation coefficient range -0.140 to -0.662). Mixed linear regression analysis between baseline logarithmic transformed ESS and neointima thickness at follow-up yielded a slope of -31 μm/ln(Pa) and a y-intercept of 99 μm. Conclusions: The hemodynamic microenvironment appears to regulate neointimal response following an Absorb BVS implantation. These findings underline the role of the ESS patterns on vessel wall healing and should be taken into consideration in the design of bioresorbable devices.
UR - http://www.scopus.com/inward/record.url?scp=84896544067&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=84896544067&partnerID=8YFLogxK
U2 - 10.1016/j.jcin.2013.05.034
DO - 10.1016/j.jcin.2013.05.034
M3 - Article
C2 - 24529931
AN - SCOPUS:84896544067
SN - 1936-8798
VL - 7
SP - 315
EP - 324
JO - JACC: Cardiovascular Interventions
JF - JACC: Cardiovascular Interventions
IS - 3
ER -