抄録

Background: We aimed to evaluate the effects of different incident directions on organ-absorbed doses, effective doses, and effective dose conversion coefficients for chest radiography. Materials and Methods: We constructed an examination environment for chest radiography using a Monte Carlo simulation and estimated the organ-absorbed doses and effective doses for frontal anterior–posterior (AP), frontal posterior–anterior (PA), left lateral, and right lateral radiography. The effective dose conversion coefficients were calculated using the air kerma-area product. Results and Discussion: Some organs, such as the breast, showed more than a fivefold difference in the organ-absorbed dose between AP and PA radiography. In lateral radiography, there was a large difference in the organ-absorbed doses in the abdominal region. The effective dose was approximately 30% higher for AP radiography than for PA radiography. The effective dose conversion coefficients differed between AP and PA radiography. These findings are consistent with the anatomical arrangement of organs in the human body; nevertheless, they have not been explicitly reported in previous studies. Conclusion: It was suggested that using the PA direction of incidence in frontal radiography reduces the effective dose received by the patient. In addition, when estimating the effective dose using conversion coefficients, the estimated value may differ significantly due to differences in the direction of incidence, so the coefficients must be appropriately selected.

本文言語英語
ページ(範囲)39-46
ページ数8
ジャーナルJournal of Radiation Protection and Research
巻51
号1
DOI
出版ステータス出版済み - 05-03-2026

UN SDG

この成果は、次の持続可能な開発目標に貢献しています

  1. SDG 3 - すべての人に健康と福祉を
    SDG 3 すべての人に健康と福祉を

All Science Journal Classification (ASJC) codes

  • 放射線
  • 放射線学、核医学およびイメージング
  • 公衆衛生学、環境および労働衛生
  • 健康、毒物学および変異誘発

フィンガープリント

「Effects of Different Incident Direction on Organ-Absorbed Doses, Effective Doses, and Effective Dose Conversion Coefficients for Chest Radiography」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル