Effects of pulsatile flow on gas transfer of membrane oxygenator: MENOX EL-4000 and Gyro C1-E3 pulsatile mode

Eiki Tayama, Yoshinari Niimi, Yoshiyuki Takami, Yukio Ohashi, Goro Ohtsuka, Tadashi Nakazawa, Julie A. Glueck, Jürgen Müller, Yukihiko Nosé

研究成果: ジャーナルへの寄稿学術論文査読

4 被引用数 (Scopus)

抄録

It is acknowledged that pulsatile flow enhances the gas exchange performance of membrane oxygenators. However, the data for currently developed oxygenators are limited. In this study, the effect of pulsatile flow was assessed utilizing the MENOX EL-4000 oxygenator. The in vitro test was performed following the Association for the Advancement of Medical Instrumentation (AAMI) standards. Pulsatile flow was produced by the Gyro C1- E3 centrifugal pump with periodical changing of the impeller speed. In Study 1, the following 3 groups were created and examined: nonpulsatile flow, pulsatile flow of 40 bpm, and pulsatile flow of 60 bpm. The blood flow rate was maintained at 3 L/min, and the V/Q ratio was 1. In Study 2, four groups were examined, nonpulsatile flow with V/Q = 1, nonpulsatile with V/Q = 2, pulsatile with V/Q = 1, and pulsatile with V/Q = 2. The blood flow rate was maintained at 4 L/min, and the pulse frequency was set at 40 bpm. In study 1, although O2 transfer was not enhanced, CO2 transfer was significantly improved (40-50%) by pulsatile flow, regardless of pulse frequency. Study 2 demonstrated that pulsatile flow resulted in improved CO2 transfer as did higher ventilation (V/Q = 2). Furthermore, even after applying higher ventilation, the pulsatile mode enhanced CO2 transfer more than the nonpulsatile mode. It was considered that the pulsatile mode induced an active secondary flow and enhanced mixing effects, and consequently CO2 transfer was improved. In conclusion, the pulsatile flow significantly enhanced the CO2 transfer of the MENOX oxygenator. It is indicated that applying the pulsatile mode is a unique and effective method to improve the gas exchange performance for a current membrane oxygenator.

本文言語英語
ページ(範囲)1127-1132
ページ数6
ジャーナルArtificial Organs
21
10
DOI
出版ステータス出版済み - 1997
外部発表はい

All Science Journal Classification (ASJC) codes

  • バイオエンジニアリング
  • 医学(その他)
  • 生体材料
  • 生体医工学

フィンガープリント

「Effects of pulsatile flow on gas transfer of membrane oxygenator: MENOX EL-4000 and Gyro C1-E3 pulsatile mode」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル