Efficacy of Intravascular Ultrasound–Based 3D Wiring Using the Tip Detection Method for CTO Intervention

Satoshi Suzuki, Atsunori Okamura, Hiroyuki Nagai, Masato Ishikawa, Satoshi Kameda, Kota Tanaka, Heitaro Watanabe, Gaku Nakazawa, Yasushi Sakata, Ichiro Shiojima

研究成果: ジャーナルへの寄稿学術論文査読

6 被引用数 (Scopus)

抄録

To perform intravascular ultrasound (IVUS)–based real-time 3-dimensional wiring in chronic total occlusion percutaneous coronary intervention, we devised a tip detection method and developed AnteOwl WR (AO)-IVUS, which is an upgraded version of Navifocus WR (Navi)-IVUS with an added pull back transducer system. We compared the procedural outcomes of AO-IVUS–based 3-dimensional wiring using the tip detection method (n = 30) and Navi-IVUS–based conventional wiring (n = 17) in chronic total occlusion percutaneous coronary intervention. The success rate of IVUS-guided wiring was markedly improved in the AO-IVUS group compared with the Navi-IVUS group (93% vs 59% of cases, respectively; P = 0.007). In cases of successful IVUS-guided wiring, the IVUS-guided wiring time was markedly improved in the AO-IVUS group compared with the Navi-IVUS group (9 ± 8 minutes vs 24 ± 26 minutes, respectively; P = 0.001). There were 2 successful cases of tip detection-antegrade dissection and re-entry in the AO-IVUS group.

本文言語英語
ページ(範囲)526-530
ページ数5
ジャーナルJACC: Asia
3
3
DOI
出版ステータス出版済み - 06-2023

All Science Journal Classification (ASJC) codes

  • 循環器および心血管医学

フィンガープリント

「Efficacy of Intravascular Ultrasound–Based 3D Wiring Using the Tip Detection Method for CTO Intervention」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル