メインナビゲーションにスキップ 検索にスキップ メインコンテンツにスキップ

Energy dependence and angular dependence of an optically stimulated luminescence dosimeter in the mammography energy range

  • Ai Kawaguchi
  • , Yuta Matsunaga
  • , Shoichi Suzuki
  • , Koichi Chida

研究成果: ジャーナルへの寄稿学術論文査読

抄録

This study aimed to investigate the energy dependence and the angular dependence of commercially available optically stimulated luminescence (OSL) point dosimeters in the mammography energy range. The energy dependence was evaluated to calculate calibration factors (CFs). The half-value layer range was 0.31-0.60 mmAl (Mo/Mo 22-28 kV, Mo/Rh 28-32 kV, and W/Rh 30-34 kV at 2-kV intervals). Mo/Rh 28 kV was the reference condition. Angular dependence was tested by rotating the X-ray tube from -90° to 90° in 30° increments, and signal counts from angled nanoDots were normalized to the 0° signal counts. Angular dependence was compared with three tube voltage and target/filter combinations (Mo/Mo 26 kV, Mo/Rh 28 kV and W/Rh 32 kV). The CFs of energy dependence were 0.94-1.06. In Mo/Mo 26-28 kV and Mo/Rh 28-32 kV, the range of CF was 0.99-1.01, which was very similar. For angular dependence, the most deteriorated normalized values (Mo/Mo, 0.37; Mo/Rh, 0.43; and W/Rh, 0.58) were observed when the X-ray tube was rotated at a 90° angle, compared to 0°. The most angular dependences of ± 30°, 60°, and 90° decreased by approximately 4%, 14%, and 63% respectively. The mean deteriorated measurement 30° intervals from 0° to ± 30° was 2%, from ± 30° to ± 60° was 8%, and from ± 60° to ± 90° was 40%. The range of energy dependence in typical mammography energy range was not as much as that in general radiography and computed tomography. For accurate measurement using nanoDot, the tilt needs to be under 30°.

本文言語英語
ページ(範囲)191-196
ページ数6
ジャーナルJournal of applied clinical medical physics
18
2
DOI
出版ステータス出版済み - 01-03-2017
外部発表はい

All Science Journal Classification (ASJC) codes

  • 放射線
  • 器械工学
  • 放射線学、核医学およびイメージング

フィンガープリント

「Energy dependence and angular dependence of an optically stimulated luminescence dosimeter in the mammography energy range」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル