T1 - Feasibility of using a pocket-sized ultrasound imaging device for detecting insulin-derived amyloidosis-like findings in patients with diabetes mellitus
T2 - a preliminary study
AU - Kageura, Naoko
AU - Takai, Hiroji
AU - Takatsu, Yasuo
AU - Suzuki, Atsushi
AU - Kobayashi, Shigeki
AU - Takano, Kazuki
AU - Nakai, Ayano
AU - Sugama, Junko
AU - Takehara, Kimie
N1 - Publisher Copyright:
Copyright © 2026 Naoko Kageura, MSN et al. This is an Open access article distributed under the Terms of Creative Commons Attribution License, which permits unrestricted use, distribution, and reproduction in any medium, provided the original author(s) and source are credited.
PY - 2026
Y1 - 2026
N2 - Objectives: Insulin-derived amyloidosis (IDA) reduces insulin absorption, increasing the risk of poor glycemic control; however, early detection remains challenging. Although ultrasound can identify IDA, the effectiveness of portable pocket-sized devices has not been evaluated. This study was performed to evaluate the detectability of findings suggestive of IDA using a pocket-sized ultrasound device on the abdominal region, the most common site for stable insulin absorption in patients with diabetes mellitus. Methods: This cross-sectional observational study was conducted in the diabetes ward of a university hospital between July and December 2024. The participants were inpatients with diabetes who had been receiving insulin injections for more than 1 month. Findings suggestive of IDA were assessed through visual inspection and palpation, magnetic resonance imaging (MRI), high-performance ultrasonography, and a pocket-sized ultrasound device. The concordance rate between the pocket-sized ultrasound device and other assessment methods was calculated. Results: Of the 20 participants enrolled, 5 met the exclusion criteria; thus, data from 15 participants were included in the final analysis. Findings suggestive of IDA were identified in nine patients (60.0%) by visual inspection and palpation and in eight patients (53.3%) by the pocket-sized ultrasound device, high-performance ultrasound, and MRI. The concordance rate among the pocket-sized ultrasound device, high-performance ultrasound, and MRI was 100%, while the concordance rate between visual inspection with palpation and the pocket-sized ultrasound device was 80%. Conclusions: These findings suggest that a pocket-sized ultrasound device has the potential to detect findings suggestive of IDA and may be suitable for use in clinical practice.
AB - Objectives: Insulin-derived amyloidosis (IDA) reduces insulin absorption, increasing the risk of poor glycemic control; however, early detection remains challenging. Although ultrasound can identify IDA, the effectiveness of portable pocket-sized devices has not been evaluated. This study was performed to evaluate the detectability of findings suggestive of IDA using a pocket-sized ultrasound device on the abdominal region, the most common site for stable insulin absorption in patients with diabetes mellitus. Methods: This cross-sectional observational study was conducted in the diabetes ward of a university hospital between July and December 2024. The participants were inpatients with diabetes who had been receiving insulin injections for more than 1 month. Findings suggestive of IDA were assessed through visual inspection and palpation, magnetic resonance imaging (MRI), high-performance ultrasonography, and a pocket-sized ultrasound device. The concordance rate between the pocket-sized ultrasound device and other assessment methods was calculated. Results: Of the 20 participants enrolled, 5 met the exclusion criteria; thus, data from 15 participants were included in the final analysis. Findings suggestive of IDA were identified in nine patients (60.0%) by visual inspection and palpation and in eight patients (53.3%) by the pocket-sized ultrasound device, high-performance ultrasound, and MRI. The concordance rate among the pocket-sized ultrasound device, high-performance ultrasound, and MRI was 100%, while the concordance rate between visual inspection with palpation and the pocket-sized ultrasound device was 80%. Conclusions: These findings suggest that a pocket-sized ultrasound device has the potential to detect findings suggestive of IDA and may be suitable for use in clinical practice.
KW - High-performance ultrasound
KW - Insulin injection
KW - Insulin-derived amyloidosis
KW - Magnetic resonance imaging
KW - Pocket-sized ultrasound
UR - https://www.scopus.com/pages/publications/105048151120
UR - https://www.scopus.com/pages/publications/105048151120#tab=citedBy
U2 - 10.20407/fmj.2025-028
DO - 10.20407/fmj.2025-028
M3 - Article
AN - SCOPUS:105048151120
SN - 2189-7247
VL - 12
SP - 143
EP - 150
JO - Fujita Medical Journal
JF - Fujita Medical Journal
IS - 2
ER -