TY - JOUR
T1 - Importance of over-reading ambulatory ECG-based microvolt T-wave alternans to eliminate three main sources of measurement error
AU - Takasugi, Nobuhiro
AU - Matsuno, Hiroko
AU - Takasugi, Mieko
AU - Shinoda, Koichi
AU - Watanabe, Takatomo
AU - Ito, Hiroyasu
AU - Okura, Hiroyuki
AU - Verrier, Richard L.
N1 - Publisher Copyright:
© 2019 Wiley Periodicals, Inc.
PY - 2019/9/1
Y1 - 2019/9/1
N2 - Background: Ambulatory electrocardiogram (ECG)-based microvolt T-wave alternans values measured by the modified moving average method (MMA-TWA) can be disrupted by T-wave changes that mimic true repolarization alternans. Methods: We investigated potential sources of measurement error by studying 19 healthy subjects (12 men; median age, 25) free of known heart disease with 36-month follow-up to establish freedom from significant arrhythmia or syncope. All participants underwent 24-hr continuous 12-lead ECG monitoring. Causes of automated MMA-TWA ≥42 µV episodes were classified based on visual inspection. Results: A total of 2,189 episodes of automated MMA-TWA episodes ≥42 µV were observed in all subjects (peak MMA-TWA: median, 94 μV; interquartile range, 81–112 μV). All episodes included one or more beats with T-wave deformation which lacked “repeating ABAB pattern” and therefore were identified as TWA measurement error. Causes of such error were categorized as: (a) artifact [72.6% (1,589/2,189), observed in 19 (100%) subjects], more frequently in limb than precordial leads; (b) T-wave changes due to changes in heart/body position [25.5% (559/2,189), observed in 14 (73.7%) subjects], frequently observed in leads V1-2; and (c) postextrasystolic T-wave changes [1.9% (41/2,189), observed in 2 (10.5%) subjects]. Conclusions: Relying only on automated MMA-TWA values obtained during ambulatory ECG monitoring can lead to incorrect measurement of TWA. Our findings offer the potential to reduce false-positive TWA results and to achieve more accurate detection of true repolarization alternans.
AB - Background: Ambulatory electrocardiogram (ECG)-based microvolt T-wave alternans values measured by the modified moving average method (MMA-TWA) can be disrupted by T-wave changes that mimic true repolarization alternans. Methods: We investigated potential sources of measurement error by studying 19 healthy subjects (12 men; median age, 25) free of known heart disease with 36-month follow-up to establish freedom from significant arrhythmia or syncope. All participants underwent 24-hr continuous 12-lead ECG monitoring. Causes of automated MMA-TWA ≥42 µV episodes were classified based on visual inspection. Results: A total of 2,189 episodes of automated MMA-TWA episodes ≥42 µV were observed in all subjects (peak MMA-TWA: median, 94 μV; interquartile range, 81–112 μV). All episodes included one or more beats with T-wave deformation which lacked “repeating ABAB pattern” and therefore were identified as TWA measurement error. Causes of such error were categorized as: (a) artifact [72.6% (1,589/2,189), observed in 19 (100%) subjects], more frequently in limb than precordial leads; (b) T-wave changes due to changes in heart/body position [25.5% (559/2,189), observed in 14 (73.7%) subjects], frequently observed in leads V1-2; and (c) postextrasystolic T-wave changes [1.9% (41/2,189), observed in 2 (10.5%) subjects]. Conclusions: Relying only on automated MMA-TWA values obtained during ambulatory ECG monitoring can lead to incorrect measurement of TWA. Our findings offer the potential to reduce false-positive TWA results and to achieve more accurate detection of true repolarization alternans.
UR - http://www.scopus.com/inward/record.url?scp=85068127887&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85068127887&partnerID=8YFLogxK
U2 - 10.1111/anec.12670
DO - 10.1111/anec.12670
M3 - Article
C2 - 31241245
AN - SCOPUS:85068127887
SN - 1082-720X
VL - 24
JO - Annals of Noninvasive Electrocardiology
JF - Annals of Noninvasive Electrocardiology
IS - 5
M1 - e12670
ER -