TY - JOUR
T1 - Improvements of lateral penumbra at various depth regions in proton pencil beam scanning with a multileaf collimator
T2 - Dose verifications and plan comparisons
AU - Tominaga, Yuki
AU - Wakisaka, Yushi
AU - Kato, Takahiro
AU - Yasui, Keisuke
AU - Kato, Ryohei
AU - Ichihara, Masaya
AU - Tomida, Masashi
AU - Sasaki, Motoharu
AU - Oita, Masataka
AU - Nishio, Teiji
N1 - Publisher Copyright:
Copyright © 2025. Published by Elsevier Ltd.
PY - 2025/12
Y1 - 2025/12
N2 - Purpose In scanned proton therapy, the current consensus is that the effective range of the collimator’s contribution to lateral penumbra improvement is up to approximately 150 mm depth. We characterized the penumbra variations for scanned proton beams with or without a new type of multileaf collimator (MLC) under various air gaps, depth, and with or without range shifter (RS). Methods Eighty-six uniform dose plans were created (38 RS-negative and 48 RS-positive plans) for nine box targets of 60 × 60 × 54 mm3 at 0–280 mm depths in water. They were created with or without MLC, with 50–300 mm air gaps. The penumbra and average doses of MLC-positive and MLC-negative plans at the organs at risk (OAR) region of each box plan were compared. Besides, several plan doses were validated by measurements with penumbra (with an average of 80–20 % dose point widths for both side profiles) differences and 2D gamma analysis. Results The MLC-positive plans reduced the penumbra and mean OAR doses by 1.0–5.1 mm and 3.3–13.5 %, respectively, compared to MLC-negative plans even at >150 mm depths. The penumbra differences in measurements were <±1.5 mm for all plans. The mean gamma scores at 2 %/2 mm were 97.9 ± 2.3 % and 97.4 ± 3.1 % for the MLC-negative and MLC-positive plans, respectively. Conclusions The MLC-positive beams improved the penumbra and reduced the OAR dose in every depth region and air gap. We have shown that PBS with MLCs can be useful at more than 150 mm regions, depending on the machine.
AB - Purpose In scanned proton therapy, the current consensus is that the effective range of the collimator’s contribution to lateral penumbra improvement is up to approximately 150 mm depth. We characterized the penumbra variations for scanned proton beams with or without a new type of multileaf collimator (MLC) under various air gaps, depth, and with or without range shifter (RS). Methods Eighty-six uniform dose plans were created (38 RS-negative and 48 RS-positive plans) for nine box targets of 60 × 60 × 54 mm3 at 0–280 mm depths in water. They were created with or without MLC, with 50–300 mm air gaps. The penumbra and average doses of MLC-positive and MLC-negative plans at the organs at risk (OAR) region of each box plan were compared. Besides, several plan doses were validated by measurements with penumbra (with an average of 80–20 % dose point widths for both side profiles) differences and 2D gamma analysis. Results The MLC-positive plans reduced the penumbra and mean OAR doses by 1.0–5.1 mm and 3.3–13.5 %, respectively, compared to MLC-negative plans even at >150 mm depths. The penumbra differences in measurements were <±1.5 mm for all plans. The mean gamma scores at 2 %/2 mm were 97.9 ± 2.3 % and 97.4 ± 3.1 % for the MLC-negative and MLC-positive plans, respectively. Conclusions The MLC-positive beams improved the penumbra and reduced the OAR dose in every depth region and air gap. We have shown that PBS with MLCs can be useful at more than 150 mm regions, depending on the machine.
KW - Lateral penumbra
KW - Multileaf collimator
KW - Pencil beam scanning
KW - Proton therapy
KW - Treatment planning
UR - https://www.scopus.com/pages/publications/105023645983
UR - https://www.scopus.com/pages/publications/105023645983#tab=citedBy
U2 - 10.1016/j.ejmp.2025.105684
DO - 10.1016/j.ejmp.2025.105684
M3 - Article
C2 - 41337947
AN - SCOPUS:105023645983
SN - 1120-1797
VL - 140
JO - Physica Medica
JF - Physica Medica
M1 - 105684
ER -