In vitro study of endotoxin adsorption by a polymyxin b-immobilized fiber column

Chizuru Yamashita, Kazuhiro Moriyama, Daisuke Hasegawa, Yuu Toki, Toshikazu Sakai, Takahiro Kawaji, Yasuyo Shimomura, Yasuyoshi Kurimoto, Mariko Nagata, Osamu Nishida

研究成果: ジャーナルへの寄稿学術論文査読

8 被引用数 (Scopus)

抄録

Background: Polymyxin B-immobilized fiber (PMX-F) columns are used as therapeutic interventions for septic shock. The clinical efficacy has been reported for 2-h applications, but their ability to adsorb endotoxin over longer treatments has not been fully elucidated. We hypothesized that PMX-F columns are capable of endotoxin removal for more than 2 h. Method: We designed closed circuits incorporating either a PMX-F column with an 8.5-mL priming volume (PMX-01R) or a sham-control column, and used inactivated fetal bovine serum as the circulating perfusate. Endotoxin was continuously injected at a fixed rate for 24 h, and perfusate endotoxin concentrations were measured at fixed time points. PMX-01R endotoxin adsorption was calculated from the difference in the endotoxin concentrations. Results: PMX-01R endotoxin adsorption increased continuously in a virtually linear manner. Conclusions: The PMX-01R column showed sustained endotoxin adsorption for at least 24 h. This indicated that PMX-F columns would be capable of clinical endotoxin removal for 24 h.

本文言語英語
ページ(範囲)269-273
ページ数5
ジャーナルBlood Purification
46
4
DOI
出版ステータス出版済み - 01-10-2018

All Science Journal Classification (ASJC) codes

  • 血液学
  • 腎臓病学

フィンガープリント

「In vitro study of endotoxin adsorption by a polymyxin b-immobilized fiber column」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル