In vivo validation of resting full-cycle ratio and diastolic pressure ratio: simultaneous measurement with instantaneous wave-free ratio

Yoshiaki Kawase, Hiroyuki Omori, Toru Tanigaki, Akihiro Hirakawa, Tetsuo Hirata, Hideaki Ota, Jun Kikuchi, Yoshihiro Sobue, Munenori Okubo, Hiroki Kamiya, Masanori Kawasaki, Takahiko Suzuki, Hitoshi Matsuo

研究成果: ジャーナルへの寄稿学術論文査読

8 被引用数 (Scopus)

抄録

The diastolic pressure ratio (dPR) and resting full-cycle ratio (RFR) are considered to be almost identical to the instantaneous wave-free ratio (iFR) in the retrospective analysis of pooled data. The aim of this study was to investigate the direct comparison of iFR and these new resting indexes in real world practice. Two pressure wires were inserted and placed in the distal part of the same coronary artery. The measurement of the iFR and the other resting indexes was performed simultaneously. A total of 54 lesions from 23 patients were subject to physiological study. In 49 lesions, iFR and other resting indexes were also measured in hyperemic conditions. The general correlation between iFR and other resting indexes was excellent in both resting and hyperemic conditions (r2 = 0.99; mean difference − 0.001 ± 0.021; p < 0.001; and r2 = 0.99; mean difference − 0.012 ± 0.025; p < 0.001, respectively). This correlation was maintained in various subgroup analyses. A diagnostic change between iFR and other resting indexes occurred in three cases (3%) when a fixed cut-off point (≤ 0.89) was applied. There was no diagnostic change when a hybrid zone (0.86 ≤ iFR ≤ 0.93) was considered. The new resting indexes and iFR showed very high correlation in real world practice. A diagnostic change only occurred in three cases (3%) when a fixed cut-off point (≤ 0.89) was applied.

本文言語英語
ページ(範囲)74-80
ページ数7
ジャーナルCardiovascular Intervention and Therapeutics
36
1
DOI
出版ステータス出版済み - 01-2021

All Science Journal Classification (ASJC) codes

  • 放射線学、核医学およびイメージング
  • 循環器および心血管医学

フィンガープリント

「In vivo validation of resting full-cycle ratio and diastolic pressure ratio: simultaneous measurement with instantaneous wave-free ratio」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル