Intervendor consistency and reproducibility of left ventricular 2D global and regional strain with two different high-end ultrasound systems

Kenji Shiino, Akira Yamada, Matthew Ischenko, Bijoy K. Khandheria, Mahala Hudaverdi, Vicki Speranza, Mary Harten, Anthony Benjamin, Christian R. Hamilton-Craig, David G. Platts, Darryl J. Burstow, Gregory M. Scalia, Jonathan Chan

研究成果: ジャーナルへの寄稿学術論文査読

44 被引用数 (Scopus)

抄録

Aims We aimed to assess intervendor agreement of global (GLS) and regional longitudinal strain by vendor-specific software after EACVI/ASE Industry Task Force Standardization Initiatives for Deformation Imaging. Methods and results Fifty-five patients underwent prospective dataset acquisitions on the same day by the same operator using two commercially available cardiac ultrasound systems (GE Vivid E9 and Philips iE33). GLS and regional peak longitudinal strain were analyzed offline using corresponding vendor-specific software (EchoPAC BT13 and QLAB version 10.3). Absolute mean GLS measurements were similar between the two vendors (GE -17.5±5.2% vs. Philips -18.9±5.1%, P = 0.15). There was excellent intervendor correlation of GLS by the same observer [r = 0.94, P<0.0001; bias 21.3%, 95% CI limits of agreement (LOA) 24.8 to 2.2%). Intervendor comparison for regional longitudinal strain by coronary artery territories distribution were: LAD: R = 0.85, P<0.0001; bias 0.5%, LOA 25.3 to 6.4%; RCA: R = 0.88, P<0.0001; bias 22.4%, LOA 28.6 to 3.7%; LCX: R = 0.76, P<0.0001; bias 25.3%, LOA 210.6 to 2.0%. Intervendor comparison for regional longitudinal strain by LV levelswere: Basal: R = 0.86, P<0.0001; bias23.6%, LOA 29.9 to 2.0%; mid: R = 0.90, P<0.0001; bias 22.6%, LOA 27.8 to 2.6%; apical: R = 0.74; P<0.0001; bias 21.3%, LOA 29.4 to 6.8%. Conclusions Intervendor agreement in GLS and regional strain measurements have significantly improved after the EACVI/ASE Task Force Strain Standardization Initiatives. However, significant wide LOA still exist, especially for regional strain measurements, which remains relevant when considering vendor-specific software for serial measurements.

本文言語英語
ページ(範囲)706-716
ページ数11
ジャーナルEuropean Heart Journal Cardiovascular Imaging
18
6
DOI
出版ステータス出版済み - 2017
外部発表はい

All Science Journal Classification (ASJC) codes

  • 放射線学、核医学およびイメージング
  • 循環器および心血管医学

フィンガープリント

「Intervendor consistency and reproducibility of left ventricular 2D global and regional strain with two different high-end ultrasound systems」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル