T1 - Neutron dose from a 6-MV X-ray beam in radiotherapy
AU - Matsubara, Hiroaki
N1 - Publisher Copyright:
© 2023, The Author(s), under exclusive licence to Japanese Society of Radiological Technology and Japan Society of Medical Physics.
PY - 2023/6
Y1 - 2023/6
N2 - Although a 6-MV X-ray beam is employed clinically as a non-neutron-producing beam, no studies have reported how few neutrons are produced from a 6-MV beam. This study aimed to theoretically deduce the neutron dose from a 6-MV beam using Monte Carlo simulations for the notification of safety and risk in radiotherapy. Nuclei from a nuclear database with neutron separation energies below 6 MeV were surveyed, suggesting that the certain content of 2H in the human body may result in some contribution. Thus, Monte Carlo calculation considering 2H in a phantom was performed. The calculation suggested that the distribution of the neutron dose from a 6-MV beam consisted of two components: one had neutrons from 2H concentrated within an irradiation field, and the other had those due to other elements such as 183W spreading from a gantry head to a treatment room. Although uncertainty owing to the normalization factor of the Monte Carlo calculations was a factor of three, the neutron doses at distances of 0 and 50 cm from an irradiation field were calculated as 27 and 1.5 nSv/MU, respectively, under intensity-modulated radiotherapy (IMRT) or volumetric modulated arc therapy (VMAT). The calculations suggest that neutrons produced by a 6-MV beam are approximately 70 and 20 times safer than those by a 10-MV beam in the case of IMRT/VMAT and total body irradiation, respectively. Thus, this study theoretically reported the approximate number of neutrons delivered by a 6-MV beam for the first time.
AB - Although a 6-MV X-ray beam is employed clinically as a non-neutron-producing beam, no studies have reported how few neutrons are produced from a 6-MV beam. This study aimed to theoretically deduce the neutron dose from a 6-MV beam using Monte Carlo simulations for the notification of safety and risk in radiotherapy. Nuclei from a nuclear database with neutron separation energies below 6 MeV were surveyed, suggesting that the certain content of 2H in the human body may result in some contribution. Thus, Monte Carlo calculation considering 2H in a phantom was performed. The calculation suggested that the distribution of the neutron dose from a 6-MV beam consisted of two components: one had neutrons from 2H concentrated within an irradiation field, and the other had those due to other elements such as 183W spreading from a gantry head to a treatment room. Although uncertainty owing to the normalization factor of the Monte Carlo calculations was a factor of three, the neutron doses at distances of 0 and 50 cm from an irradiation field were calculated as 27 and 1.5 nSv/MU, respectively, under intensity-modulated radiotherapy (IMRT) or volumetric modulated arc therapy (VMAT). The calculations suggest that neutrons produced by a 6-MV beam are approximately 70 and 20 times safer than those by a 10-MV beam in the case of IMRT/VMAT and total body irradiation, respectively. Thus, this study theoretically reported the approximate number of neutrons delivered by a 6-MV beam for the first time.
UR - http://www.scopus.com/inward/record.url?scp=85147936894&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85147936894&partnerID=8YFLogxK
U2 - 10.1007/s12194-023-00705-6
DO - 10.1007/s12194-023-00705-6
M3 - Article
C2 - 36780121
AN - SCOPUS:85147936894
SN - 1865-0333
VL - 16
SP - 186
EP - 194
JO - Radiological Physics and Technology
JF - Radiological Physics and Technology
IS - 2
ER -