抄録
This study presents a comparison of novel pregnant model phantoms with a handmade phantom in terms of shape and radiation measurement points to determine which model is more suitable for measuring the foetal radiation dose during x-ray examinations. Novel pregnant model phantoms were constructed using an anthropomorphic phantom in combination with two differently-sized custom-made abdomen phantoms simulating pregnancy, which were constructed from a polyurethane resin. The size and shape of the polyurethane resin were designed based on abdominal sizes and shapes collected from the computed tomography examinations at 18 pregnant patients of one hospital. The handmade pregnant model phantom was constructed using an anthropomorphic phantom and a beach ball containing water. Compared with the handmade phantom, there were additional dose measurement points on the novel pregnant model phantoms. Our model phantoms improved upon the handmade phantom in terms of shape and radiation measurement points. We produced pregnant model phantoms that simulated the shapes and sizes of actual patients for the first time.
| 本文言語 | 英語 |
|---|---|
| ページ(範囲) | N12-N21 |
| ジャーナル | Journal of Radiological Protection |
| 巻 | 41 |
| 号 | 3 |
| DOI | |
| 出版ステータス | 出版済み - 09-2021 |
UN SDG
この成果は、次の持続可能な開発目標に貢献しています
-
SDG 3 すべての人に健康と福祉を
All Science Journal Classification (ASJC) codes
- 廃棄物管理と処理
- 公衆衛生学、環境および労働衛生
フィンガープリント
「Novel pregnant model phantoms for measurement of foetal radiation dose in x-ray examinations」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。引用スタイル
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver