抄録
We objectively evaluate a straightforward registration method for correcting respiration-induced movement of abdominal organs in CT perfusion studies by measuring the distributions of alignment errors between corresponding landmark pairs. We introduce the concept and describe the advantages of using the surface-normal component of distance between pairs of corresponding landmarks selected so that their surface normal is in one of the three coordinate axis directions, and show that such landmarks can be precisely placed with respect to the surface normal. Using a large population of landmark pairs on a substantial quantity of 4D dynamic contrast-enhanced CT volume data, we quantify the average alignment errors of abdominal organs that remain uncorrected by registration.
| 本文言語 | 英語 |
|---|---|
| ページ(範囲) | 1701-1715 |
| ページ数 | 15 |
| ジャーナル | Physics in Medicine and Biology |
| 巻 | 57 |
| 号 | 6 |
| DOI | |
| 出版ステータス | 出版済み - 21-03-2012 |
| 外部発表 | はい |
All Science Journal Classification (ASJC) codes
- 放射線技術および超音波技術
- 放射線学、核医学およびイメージング
フィンガープリント
「Objective evaluation of the correction by non-rigid registration of abdominal organ motion in low-dose 4D dynamic contrast-enhanced CT」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。引用スタイル
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver