Patient investigation of average glandular dose and incident air kerma for digital mammography

  • Ai Kawaguchi
  • , Yuta Matsunaga
  • , Tomoko Otsuka
  • , Shoichi Suzuki

研究成果: ジャーナルへの寄稿学術論文査読

抄録

Our aim in this study was to investigate the incident air kerma (IAK) and average glandular dose (AGD) for polymethylmethacrylate (PMMA) phantoms and patient breasts for each thickness by use of digital mammography units, and to determine the correlation between phantom and patient measurements. An additional aim was to calculate the numerical value of the diagnostic reference level (DRL) for digital mammography in Japan based on the AGD from patient measurements. Patient-based IAK and AGD values were calculated for 300 patients who underwent mammographic examinations at three institutions. On examination of a 40-mm PMMA phantom, the IAK and the AGD were 7.89 and 1.84 mGy, respectively. The mean patient breast thickness was 37.6 mm, and the mean IAK and the AGD calculated from actual patient data for breast thicknesses between 40 mm and 50 mm were 8.91 and 2.08 mGy, respectively. Approximately 20% of the 300 patients had IAK >10 mGy. The distributions of patient-based IAK and AGD values were higher than the IAK and AGD values for the PMMA phantom. The patient dose with use of the PMMA phantom can be underestimated. The DRL was calculated from patient-based AGDs of the regular breast thickness as 2.0 mGy in Japan.

本文言語英語
ページ(範囲)102-108
ページ数7
ジャーナルRadiological Physics and Technology
7
1
DOI
出版ステータス出版済み - 01-2014
外部発表はい

All Science Journal Classification (ASJC) codes

  • 放射線
  • 理学療法、スポーツ療法とリハビリテーション
  • 放射線学、核医学およびイメージング

フィンガープリント

「Patient investigation of average glandular dose and incident air kerma for digital mammography」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル