[Physical image properties of digital radiography systems in low dose range].

Hiroshi Kunitomo, Katsuhiro Ichikawa, Ryo Higashide, Kazuya Ohashi

研究成果: ジャーナルへの寄稿学術論文査読

1 被引用数 (Scopus)

抄録

We measured physical image properties of a flat panel detector (FPD) system and a computed radiography (CR) system, targeting to a low dose range (reference dose: 2.58×10(-7) C/kg: to 1/20 dose). Input-output properties, pre-sampled modulation transfer functions (pre-sampled MTFs), and normalized noise power spectra for an FPD system equipped with a CsI scintillator (FPDcsi) and a CR system with an imaging plate coated with storage phosphor (CR) were measured at the low dose range for radiation quality of RQA3 (50 skV) and RQA5 (70 kV), and detective quantum efficiencies (DQEs) were calculated. In addition, in order to validate the DQE results, component fractions of Poisson and multiplicative and additive noise were analyzed using relative standard deviation analysis. The DQE values of FPDcsi were decreased with dose decrease, and contrarily to these, those of CR were increased. At the 1/10 and 1/20 doses for RQA3, the DQEs of FPDcsi and CR became almost the same. From the results of RSD analysis, it was proved that the main cause of DQE deterioration on FPDcsi are non-negligible additive (electronic) noise, and the DQE improvement on CR was caused by both of significant multiplicative (structure) noise and very low electronic noise. The DQE results were validated by comparing burger phantom images of each dose and radiation quality.

本文言語英語
ページ(範囲)961-969
ページ数9
ジャーナルNihon Hoshasen Gijutsu Gakkai zasshi
68
8
DOI
出版ステータス出版済み - 2012
外部発表はい

All Science Journal Classification (ASJC) codes

  • 医学一般

フィンガープリント

「[Physical image properties of digital radiography systems in low dose range].」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル